DEVELOPMENT OF POLY (MELAMINE) MODIFIED CARBON PASTE ELECTRODE AS SENSOR ASCORBIC ACID, URIC ACID AND ACETAMINOPHEN

Melamin is s a toxic compound that is used in the manufacture of plastic tableware and white glue. However, poly(melamine) can be used to modified carbon paste electrode (CPE) as working electrode in voltammetry. In this research, film of melamine was prepared on the surface of CPE through electropo...

全面介紹

Saved in:
書目詳細資料
主要作者: KUSDIANTINI (NIM : 20514011), ANDINI
格式: Theses
語言:Indonesia
在線閱讀:https://digilib.itb.ac.id/gdl/view/19878
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:Melamin is s a toxic compound that is used in the manufacture of plastic tableware and white glue. However, poly(melamine) can be used to modified carbon paste electrode (CPE) as working electrode in voltammetry. In this research, film of melamine was prepared on the surface of CPE through electropolymerisation. Electropolymerisation was conducted using cyclic voltammetry (CV) technique (20 cycle) at potensial range of 0 V – 1,6 V and scan rate of 100 mV/sfrom solution containing 1 mM melamine and 0.1 M H2SO4. The modified electrode was used to determined ascorbic acid, acetaminophen and uric acid. Linear calibration curve was found at concentration range of 79 µM - 18 mM for ascorbic acid,4 µM - 1,14 mM for acetaminophene, and 10 µM - 2 mM for uric acid. The limit of detection forascorbic acid, acetaminophen, and uric acid, was 47 µM ,1,68 µM and 22 µM, respectively. Results of voltammetric determination of ascorbic acid in real samples are in accordance with those of spectrophotometric methode.