Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I

Point-of-care testing (POCT) of cardiac troponin device is aimed for improvement in turn round time (TAT) and assist in acute management care of acute coronary syndrome (ACS). The present study was conducted to assess the analytical performance and correlation of HUBI-QUANPro troponin I with an ex...

Full description

Saved in:
Bibliographic Details
Main Authors: Munirah Mansor, Mohd Nazrul AA, Norsheila M, Dian Nasriana N, Hanita O
Format: Article
Language:English
Published: Pusat Perubatan Universiti Kebangsaan Malaysia 2019
Online Access:http://journalarticle.ukm.my/15549/1/7_ms0300_pdf_10498.pdf
http://journalarticle.ukm.my/15549/
https://www.medicineandhealthukm.com/toc/14/2
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Kebangsaan Malaysia
Language: English
id my-ukm.journal.15549
record_format eprints
spelling my-ukm.journal.155492020-11-04T15:40:58Z http://journalarticle.ukm.my/15549/ Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I Munirah Mansor, Mohd Nazrul AA, Norsheila M, Dian Nasriana N, Hanita O, Point-of-care testing (POCT) of cardiac troponin device is aimed for improvement in turn round time (TAT) and assist in acute management care of acute coronary syndrome (ACS). The present study was conducted to assess the analytical performance and correlation of HUBI-QUANPro troponin I with an existing laboratory instrument of high-sensitivity troponin I, Abbott Architect. The factors that were studied, included precision study by using manufacturer quality control (QC) material (2 levels) and correlation study of sample differences (whole blood, plasma and serum) and methodology (immunochromatographic assay and chemiluminescent immunoassay). A total of 30 QC was used for precision study and 42 sample serum and EDTA for the correlation study. An acceptable total imprecision of 10.9% and 6.7% were seen at level of 0.91 ng/mL and 2.66 ng/mL, respectively. Regression analysis of sample differences (plasma vs whole blood) in HUBI-QUANPro showed slope of 0.935, r=0.991 (p=<0.001). Correlation of HUBI-QUANPro and Abbott Architect (whole blood, plasma vs serum) both demonstrated regression slope of 0.205, r=0.963 (whole blood) and slope of 0.192, r=0.954 (plasma), p=<0.001, respectively. HUBI-QUANPro troponin I POCT device is a sensitive, fast, precise and has a good comparable analytical performance with reference laboratory instrument for cardiac troponin I measurement. It is able to serve as a good POCT device in cardiac-related acute care management. Pusat Perubatan Universiti Kebangsaan Malaysia 2019 Article PeerReviewed application/pdf en http://journalarticle.ukm.my/15549/1/7_ms0300_pdf_10498.pdf Munirah Mansor, and Mohd Nazrul AA, and Norsheila M, and Dian Nasriana N, and Hanita O, (2019) Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I. Medicine & Health, 14 (2). pp. 77-86. ISSN 2289-5728 https://www.medicineandhealthukm.com/toc/14/2
institution Universiti Kebangsaan Malaysia
building Tun Sri Lanang Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Kebangsaan Malaysia
content_source UKM Journal Article Repository
url_provider http://journalarticle.ukm.my/
language English
description Point-of-care testing (POCT) of cardiac troponin device is aimed for improvement in turn round time (TAT) and assist in acute management care of acute coronary syndrome (ACS). The present study was conducted to assess the analytical performance and correlation of HUBI-QUANPro troponin I with an existing laboratory instrument of high-sensitivity troponin I, Abbott Architect. The factors that were studied, included precision study by using manufacturer quality control (QC) material (2 levels) and correlation study of sample differences (whole blood, plasma and serum) and methodology (immunochromatographic assay and chemiluminescent immunoassay). A total of 30 QC was used for precision study and 42 sample serum and EDTA for the correlation study. An acceptable total imprecision of 10.9% and 6.7% were seen at level of 0.91 ng/mL and 2.66 ng/mL, respectively. Regression analysis of sample differences (plasma vs whole blood) in HUBI-QUANPro showed slope of 0.935, r=0.991 (p=<0.001). Correlation of HUBI-QUANPro and Abbott Architect (whole blood, plasma vs serum) both demonstrated regression slope of 0.205, r=0.963 (whole blood) and slope of 0.192, r=0.954 (plasma), p=<0.001, respectively. HUBI-QUANPro troponin I POCT device is a sensitive, fast, precise and has a good comparable analytical performance with reference laboratory instrument for cardiac troponin I measurement. It is able to serve as a good POCT device in cardiac-related acute care management.
format Article
author Munirah Mansor,
Mohd Nazrul AA,
Norsheila M,
Dian Nasriana N,
Hanita O,
spellingShingle Munirah Mansor,
Mohd Nazrul AA,
Norsheila M,
Dian Nasriana N,
Hanita O,
Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
author_facet Munirah Mansor,
Mohd Nazrul AA,
Norsheila M,
Dian Nasriana N,
Hanita O,
author_sort Munirah Mansor,
title Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
title_short Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
title_full Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
title_fullStr Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
title_full_unstemmed Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
title_sort analytical evaluation of cardiac poct: humasis hubi-quanpro troponin i
publisher Pusat Perubatan Universiti Kebangsaan Malaysia
publishDate 2019
url http://journalarticle.ukm.my/15549/1/7_ms0300_pdf_10498.pdf
http://journalarticle.ukm.my/15549/
https://www.medicineandhealthukm.com/toc/14/2
_version_ 1683231759607529472