Box–Behnken design based development of UV-reversed phase high performance liquid chromatographic method for determination of ascorbic acid in tablet formulations
A simple, sensitive, accurate and inexpensive UV-reversed-phase high-performance liquid chromatographic method was developed for the determination of ascorbic acid in tablet formulations. The method was based on the separation of ascorbic acid using a mobile phase of an acetonitrileNaH2PO4 -H3PO4...
Saved in:
Main Authors: | , , , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI
2022
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/101225/7/101225_Box%E2%80%93Behnken%20design%20based%20development.pdf http://irep.iium.edu.my/101225/ https://www.mdpi.com/2297-8739/9/11/361/htm |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Islam Antarabangsa Malaysia |
Language: | English |
Summary: | A simple, sensitive, accurate and inexpensive UV-reversed-phase high-performance liquid
chromatographic method was developed for the determination of ascorbic acid in tablet formulations.
The method was based on the separation of ascorbic acid using a mobile phase of an acetonitrileNaH2PO4
-H3PO4 buffer solution (pH = 3) (5:95 v/v) with a UV detection wavelength of 245 nm
and a flow rate of 0.8 mL min−1 at ambient column temperature. The variables of the proposed
method, such as acetonitrile fraction (%), flow rate (mL min−1
) and column temperature (◦C), were
optimized on the peak area by response surface methodology via the Box–Behnken design. The
mobile phase was passed isocratically, and the separation of ascorbic acid was performed at the
retention time of 4.1 min. A calibration graph was obtained and found to be linear in the concentration
range of 10–180 µg mL−1
. The method suitability was assessed and an asymmetry factor of 1.15
was obtained. The proposed method was successfully applied for the determination of ascorbic
acid in tablet formulations and statistically compared with the results of the reference method. The
performance of the proposed method was excellent and in agreement with the reference method. The
recovery percentage of the proposed and reference methods was in the range of 99.98–100.04% and
showed compliance (100 ± 2%) with regulatory guidelines. |
---|