Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer on the same day and 3 months apart. Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements made on the same day (n =...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
American Academy of Optometry
2015
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/46135/1/OVS_2015.pdf http://irep.iium.edu.my/46135/ http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Islam Antarabangsa Malaysia |
Language: | English |
id |
my.iium.irep.46135 |
---|---|
record_format |
dspace |
spelling |
my.iium.irep.461352015-12-21T01:53:23Z http://irep.iium.edu.my/46135/ Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer Chao, Cecilia Stapleton, Fiona Badarudin, Noor Ezailina Golebiowski, Blanka RE Ophthalmology Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer on the same day and 3 months apart. Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements made on the same day (n = 20 subjects) and 3 months apart (n = 29 subjects). The Cochet-Bonnet esthesiometer was used to measure corneal and inferior conjunctival thresholds using the ascending method of limits. The pressure exerted by the Cochet-Bonnet esthesiometer was determined using an analytical balance, for both the 0.08- and 0.12-mm-diameter filaments. This calibration was then used to convert filament length measurements to pressure. Repeatability was determined using a Bland and Altman analysis. Results. The pressure exerted at each filament length differed between the two filament diameters. The measured pressure also differed from values provided by the manufacturer. Repeatability of threshold measurements at the central cornea was shown to be good, with better repeatability for same-day measurements (coefficient of repeatability [CoR] = T0.23 g/mm2) than for those 3 months apart (CoR = T0.52 g/mm2). Threshold measurements at the inferior conjunctiva, in contrast, were poorly repeatable (CoR = T12.78 g/mm2). Conclusions. Cochet-Bonnet esthesiometry is repeatable when performed on the central cornea on the same day and 3 months apart, but this instrument is not recommended for conjunctival threshold measurements. (Optom Vis Sci 2015;92:183Y189) Key Words: Cochet-Bonnet esthesiometer, repeatability, ocular surface sensitivity, corneal mechanical threshold, conjunctival mechanical threshold American Academy of Optometry 2015-02 Article REM application/pdf en http://irep.iium.edu.my/46135/1/OVS_2015.pdf Chao, Cecilia and Stapleton, Fiona and Badarudin, Noor Ezailina and Golebiowski, Blanka (2015) Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer. Optometry and Vision Science, 92 (2). pp. 183-189. ISSN 1538-9235 (O), 1040-5488 (P) http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx 10.1097/OPX.0000000000000472 |
institution |
Universiti Islam Antarabangsa Malaysia |
building |
IIUM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
International Islamic University Malaysia |
content_source |
IIUM Repository (IREP) |
url_provider |
http://irep.iium.edu.my/ |
language |
English |
topic |
RE Ophthalmology |
spellingShingle |
RE Ophthalmology Chao, Cecilia Stapleton, Fiona Badarudin, Noor Ezailina Golebiowski, Blanka Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
description |
Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer
on the same day and 3 months apart.
Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements
made on the same day (n = 20 subjects) and 3 months apart (n = 29 subjects). The Cochet-Bonnet esthesiometer was used to
measure corneal and inferior conjunctival thresholds using the ascending method of limits. The pressure exerted by the
Cochet-Bonnet esthesiometer was determined using an analytical balance, for both the 0.08- and 0.12-mm-diameter filaments.
This calibration was then used to convert filament length measurements to pressure. Repeatability was determined
using a Bland and Altman analysis.
Results. The pressure exerted at each filament length differed between the two filament diameters. The measured pressure
also differed from values provided by the manufacturer. Repeatability of threshold measurements at the central cornea was
shown to be good, with better repeatability for same-day measurements (coefficient of repeatability [CoR] = T0.23 g/mm2)
than for those 3 months apart (CoR = T0.52 g/mm2). Threshold measurements at the inferior conjunctiva, in contrast, were
poorly repeatable (CoR = T12.78 g/mm2).
Conclusions. Cochet-Bonnet esthesiometry is repeatable when performed on the central cornea on the same day and
3 months apart, but this instrument is not recommended for conjunctival threshold measurements.
(Optom Vis Sci 2015;92:183Y189)
Key Words: Cochet-Bonnet esthesiometer, repeatability, ocular surface sensitivity, corneal mechanical threshold,
conjunctival mechanical threshold |
format |
Article |
author |
Chao, Cecilia Stapleton, Fiona Badarudin, Noor Ezailina Golebiowski, Blanka |
author_facet |
Chao, Cecilia Stapleton, Fiona Badarudin, Noor Ezailina Golebiowski, Blanka |
author_sort |
Chao, Cecilia |
title |
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
title_short |
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
title_full |
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
title_fullStr |
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
title_full_unstemmed |
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
title_sort |
ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer |
publisher |
American Academy of Optometry |
publishDate |
2015 |
url |
http://irep.iium.edu.my/46135/1/OVS_2015.pdf http://irep.iium.edu.my/46135/ http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx |
_version_ |
1643612931412197376 |