Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer

Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer on the same day and 3 months apart. Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements made on the same day (n =...

Full description

Saved in:
Bibliographic Details
Main Authors: Chao, Cecilia, Stapleton, Fiona, Badarudin, Noor Ezailina, Golebiowski, Blanka
Format: Article
Language:English
Published: American Academy of Optometry 2015
Subjects:
Online Access:http://irep.iium.edu.my/46135/1/OVS_2015.pdf
http://irep.iium.edu.my/46135/
http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Islam Antarabangsa Malaysia
Language: English
id my.iium.irep.46135
record_format dspace
spelling my.iium.irep.461352015-12-21T01:53:23Z http://irep.iium.edu.my/46135/ Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer Chao, Cecilia Stapleton, Fiona Badarudin, Noor Ezailina Golebiowski, Blanka RE Ophthalmology Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer on the same day and 3 months apart. Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements made on the same day (n = 20 subjects) and 3 months apart (n = 29 subjects). The Cochet-Bonnet esthesiometer was used to measure corneal and inferior conjunctival thresholds using the ascending method of limits. The pressure exerted by the Cochet-Bonnet esthesiometer was determined using an analytical balance, for both the 0.08- and 0.12-mm-diameter filaments. This calibration was then used to convert filament length measurements to pressure. Repeatability was determined using a Bland and Altman analysis. Results. The pressure exerted at each filament length differed between the two filament diameters. The measured pressure also differed from values provided by the manufacturer. Repeatability of threshold measurements at the central cornea was shown to be good, with better repeatability for same-day measurements (coefficient of repeatability [CoR] = T0.23 g/mm2) than for those 3 months apart (CoR = T0.52 g/mm2). Threshold measurements at the inferior conjunctiva, in contrast, were poorly repeatable (CoR = T12.78 g/mm2). Conclusions. Cochet-Bonnet esthesiometry is repeatable when performed on the central cornea on the same day and 3 months apart, but this instrument is not recommended for conjunctival threshold measurements. (Optom Vis Sci 2015;92:183Y189) Key Words: Cochet-Bonnet esthesiometer, repeatability, ocular surface sensitivity, corneal mechanical threshold, conjunctival mechanical threshold American Academy of Optometry 2015-02 Article REM application/pdf en http://irep.iium.edu.my/46135/1/OVS_2015.pdf Chao, Cecilia and Stapleton, Fiona and Badarudin, Noor Ezailina and Golebiowski, Blanka (2015) Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer. Optometry and Vision Science, 92 (2). pp. 183-189. ISSN 1538-9235 (O), 1040-5488 (P) http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx 10.1097/OPX.0000000000000472
institution Universiti Islam Antarabangsa Malaysia
building IIUM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider International Islamic University Malaysia
content_source IIUM Repository (IREP)
url_provider http://irep.iium.edu.my/
language English
topic RE Ophthalmology
spellingShingle RE Ophthalmology
Chao, Cecilia
Stapleton, Fiona
Badarudin, Noor Ezailina
Golebiowski, Blanka
Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
description Purpose. To determine the repeatability of ocular surface threshold measurements using the Cochet-Bonnet esthesiometer on the same day and 3 months apart. Methods. Two separate studies were conducted to determine the repeatability of ocular surface threshold measurements made on the same day (n = 20 subjects) and 3 months apart (n = 29 subjects). The Cochet-Bonnet esthesiometer was used to measure corneal and inferior conjunctival thresholds using the ascending method of limits. The pressure exerted by the Cochet-Bonnet esthesiometer was determined using an analytical balance, for both the 0.08- and 0.12-mm-diameter filaments. This calibration was then used to convert filament length measurements to pressure. Repeatability was determined using a Bland and Altman analysis. Results. The pressure exerted at each filament length differed between the two filament diameters. The measured pressure also differed from values provided by the manufacturer. Repeatability of threshold measurements at the central cornea was shown to be good, with better repeatability for same-day measurements (coefficient of repeatability [CoR] = T0.23 g/mm2) than for those 3 months apart (CoR = T0.52 g/mm2). Threshold measurements at the inferior conjunctiva, in contrast, were poorly repeatable (CoR = T12.78 g/mm2). Conclusions. Cochet-Bonnet esthesiometry is repeatable when performed on the central cornea on the same day and 3 months apart, but this instrument is not recommended for conjunctival threshold measurements. (Optom Vis Sci 2015;92:183Y189) Key Words: Cochet-Bonnet esthesiometer, repeatability, ocular surface sensitivity, corneal mechanical threshold, conjunctival mechanical threshold
format Article
author Chao, Cecilia
Stapleton, Fiona
Badarudin, Noor Ezailina
Golebiowski, Blanka
author_facet Chao, Cecilia
Stapleton, Fiona
Badarudin, Noor Ezailina
Golebiowski, Blanka
author_sort Chao, Cecilia
title Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
title_short Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
title_full Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
title_fullStr Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
title_full_unstemmed Ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
title_sort ocular surface sensitivity repeatability with cochet-bonnet aesthesiometer
publisher American Academy of Optometry
publishDate 2015
url http://irep.iium.edu.my/46135/1/OVS_2015.pdf
http://irep.iium.edu.my/46135/
http://journals.lww.com/optvissci/Abstract/2015/02000/Ocular_Surface_Sensitivity_Repeatability_with.11.aspx
_version_ 1643612931412197376