An alternate voltage-controlled current source for electrical impedance tomography applications
The desirable features of any current source in electrical impedance tomography (EIT) are high output impedance and a constant current for varying loads. Howland current sources are the most popular current sources for EIT applications. Alternately, a simple non-inverting amplifier circuit can be us...
Saved in:
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Book Chapter |
Published: |
Springer
2023
|
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Tenaga Nasional |
Summary: | The desirable features of any current source in electrical impedance tomography (EIT) are high output impedance and a constant current for varying loads. Howland current sources are the most popular current sources for EIT applications. Alternately, a simple non-inverting amplifier circuit can be used as a voltage-controlled current source using IC LM7171 op-amp. A constant current of 5 mA (PP) is designed and tested over a wide range of frequencies. The frequency response indicates a stable current up to 1 MHz. These results are validated for an impedance of 1 k? which is usually considered for EIT applications. A buffer circuit is added to this current circuit to meet high output impedance matching. A minor discrepancy is visible between the theoretical value and measured value of the output current due to the tolerance band of the resistors used in the non-inverting amplifier circuit. � Springer Nature Singapore Pte Ltd 2019. |
---|