Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis

Long exposure to radiation from fluoroscopy-guided interventions (FGIs) can be detrimental to both patients and radiologists. The effective doses received by the interventional radiology staff after performing 230 FGIs in a year were assessed by using double dosimetry and five various algorithms. Th...

Full description

Saved in:
Bibliographic Details
Main Authors: Bohari, Asmah, Hashim, Suhairul, Ghoshal, Sib Krishna, Mohd Mustafa, Siti Norsyafiqah
Format: Article
Published: Oxford University Press 2019
Subjects:
Online Access:http://eprints.utm.my/id/eprint/88314/
https://doi.org/10.1093/rpd/ncz051
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Teknologi Malaysia
id my.utm.88314
record_format eprints
spelling my.utm.883142020-12-14T23:58:37Z http://eprints.utm.my/id/eprint/88314/ Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis Bohari, Asmah Hashim, Suhairul Ghoshal, Sib Krishna Mohd Mustafa, Siti Norsyafiqah QC Physics Long exposure to radiation from fluoroscopy-guided interventions (FGIs) can be detrimental to both patients and radiologists. The effective doses received by the interventional radiology staff after performing 230 FGIs in a year were assessed by using double dosimetry and five various algorithms. The Shapiro–Wilk test revealed normally-distributed data (p < 0.01), while the significant correlation coefficients between the effective doses ranged between 0.88 and 1.00. As for the Bland–Altman analysis, both Niklason and Boetticher algorithms strongly supported the absence of statistical significance between the estimated effective doses. This portrays that the occupational doses received by the interventional radiology staff during FGIs fall within the acceptable limit regardless of the varied algorithms applied. In short, the Niklason and Boetticher algorithms appeared to be the more interchangeable ones for effective evaluation of doses. This is in view of their strong mutual correlations and excellent agreement. Oxford University Press 2019 Article PeerReviewed Bohari, Asmah and Hashim, Suhairul and Ghoshal, Sib Krishna and Mohd Mustafa, Siti Norsyafiqah (2019) Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis. Radiation Protection Dosimetry, 186 (4). pp. 462-468. ISSN 1742-3406 https://doi.org/10.1093/rpd/ncz051 DOI:10.1093/rpd/ncz051
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
topic QC Physics
spellingShingle QC Physics
Bohari, Asmah
Hashim, Suhairul
Ghoshal, Sib Krishna
Mohd Mustafa, Siti Norsyafiqah
Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
description Long exposure to radiation from fluoroscopy-guided interventions (FGIs) can be detrimental to both patients and radiologists. The effective doses received by the interventional radiology staff after performing 230 FGIs in a year were assessed by using double dosimetry and five various algorithms. The Shapiro–Wilk test revealed normally-distributed data (p < 0.01), while the significant correlation coefficients between the effective doses ranged between 0.88 and 1.00. As for the Bland–Altman analysis, both Niklason and Boetticher algorithms strongly supported the absence of statistical significance between the estimated effective doses. This portrays that the occupational doses received by the interventional radiology staff during FGIs fall within the acceptable limit regardless of the varied algorithms applied. In short, the Niklason and Boetticher algorithms appeared to be the more interchangeable ones for effective evaluation of doses. This is in view of their strong mutual correlations and excellent agreement.
format Article
author Bohari, Asmah
Hashim, Suhairul
Ghoshal, Sib Krishna
Mohd Mustafa, Siti Norsyafiqah
author_facet Bohari, Asmah
Hashim, Suhairul
Ghoshal, Sib Krishna
Mohd Mustafa, Siti Norsyafiqah
author_sort Bohari, Asmah
title Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
title_short Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
title_full Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
title_fullStr Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
title_full_unstemmed Assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using Bland-Altman analysis
title_sort assessment on the interchangeability of personal effective dose algorithms in fluoroscopy-guided interventions using bland-altman analysis
publisher Oxford University Press
publishDate 2019
url http://eprints.utm.my/id/eprint/88314/
https://doi.org/10.1093/rpd/ncz051