Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia

The utilization and prolonged working of two gamma spectrometry (GS1 and GS2) in nuclear laboratories, UTM, causing one to questioning its performance. To achievethe higher quality outcomes of gamma spectrometry system. Itsperformance specifications should verify against the warranted values offere...

Full description

Saved in:
Bibliographic Details
Main Authors: Usman, Abubakar, MohdNor, Norehan
Format: Article
Language:English
Published: International Organization of Scientific Research (IOSR) 2019
Subjects:
Online Access:http://eprints.utm.my/id/eprint/88558/1/NorehanMohdNor2019_TestperformanceofGammaSpectrometryCoAxial.pdf
http://eprints.utm.my/id/eprint/88558/
https://www.iosrjournals.org/iosr-jap/papers/Vol11-issue6/Series-1/I1106015866.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Teknologi Malaysia
Language: English
id my.utm.88558
record_format eprints
spelling my.utm.885582020-12-15T10:53:25Z http://eprints.utm.my/id/eprint/88558/ Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia Usman, Abubakar MohdNor, Norehan Q Science (General) The utilization and prolonged working of two gamma spectrometry (GS1 and GS2) in nuclear laboratories, UTM, causing one to questioning its performance. To achievethe higher quality outcomes of gamma spectrometry system. Itsperformance specifications should verify against the warranted values offered by the manufacturer. High purity germanium (HPGe) detectors is the most distinguished radiation measurement instrument that produced excellent energy resolution. The aims of this studyis to determine the working condition and compare the performances of two gamma spectrometry systems. The GS1 consist of n-type closed end coaxial HPGe detector GC 2018 model and GS2 consist of p-type closed end coaxial HPGe detector of GEM25-76-LB-C model. The test performance specifications such as resolution, peak shapes, peak-to-Compton ratio, figure of merit, and dead time for both spectrometry systems are measured using American ANSI/IEEE 325-1996 standard procedure.Four (4) standard source 60Co, 152Eu, 133Ba, and 137Cs were used. It covered energies range from (4.3 keV-3194.9) keV. The source-to-detector distance is set 25 cm to avoid the summing coincident gamma ray. The relative efficiency measured improve by 0.2% and 4% for GS1 and GS2 respectively. Peak-to-Compton ratio of both detectors improved by factor of 4.Dead time found to be less than 1% at 25 cm compared to 12 cm. From the results, GS1 has higher resolution compared to GS2 detector. Based on the results obtained, it can be concluded that the performance of two coaxial HPGe detectors in nuclear laboratories, (UTM) are in good working condition.Thisrevealed propercontrol and maintenance of the two detectors. International Organization of Scientific Research (IOSR) 2019 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/88558/1/NorehanMohdNor2019_TestperformanceofGammaSpectrometryCoAxial.pdf Usman, Abubakar and MohdNor, Norehan (2019) Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia. IOSR Journal of Applied Physics, 11 (6). pp. 58-66. ISSN 2278-4861 https://www.iosrjournals.org/iosr-jap/papers/Vol11-issue6/Series-1/I1106015866.pdf
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic Q Science (General)
spellingShingle Q Science (General)
Usman, Abubakar
MohdNor, Norehan
Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
description The utilization and prolonged working of two gamma spectrometry (GS1 and GS2) in nuclear laboratories, UTM, causing one to questioning its performance. To achievethe higher quality outcomes of gamma spectrometry system. Itsperformance specifications should verify against the warranted values offered by the manufacturer. High purity germanium (HPGe) detectors is the most distinguished radiation measurement instrument that produced excellent energy resolution. The aims of this studyis to determine the working condition and compare the performances of two gamma spectrometry systems. The GS1 consist of n-type closed end coaxial HPGe detector GC 2018 model and GS2 consist of p-type closed end coaxial HPGe detector of GEM25-76-LB-C model. The test performance specifications such as resolution, peak shapes, peak-to-Compton ratio, figure of merit, and dead time for both spectrometry systems are measured using American ANSI/IEEE 325-1996 standard procedure.Four (4) standard source 60Co, 152Eu, 133Ba, and 137Cs were used. It covered energies range from (4.3 keV-3194.9) keV. The source-to-detector distance is set 25 cm to avoid the summing coincident gamma ray. The relative efficiency measured improve by 0.2% and 4% for GS1 and GS2 respectively. Peak-to-Compton ratio of both detectors improved by factor of 4.Dead time found to be less than 1% at 25 cm compared to 12 cm. From the results, GS1 has higher resolution compared to GS2 detector. Based on the results obtained, it can be concluded that the performance of two coaxial HPGe detectors in nuclear laboratories, (UTM) are in good working condition.Thisrevealed propercontrol and maintenance of the two detectors.
format Article
author Usman, Abubakar
MohdNor, Norehan
author_facet Usman, Abubakar
MohdNor, Norehan
author_sort Usman, Abubakar
title Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
title_short Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
title_full Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
title_fullStr Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
title_full_unstemmed Test performance of Gamma Spectrometry Co-Axial High Purity Germanium detectorsin Universiti Teknologi Malaysia
title_sort test performance of gamma spectrometry co-axial high purity germanium detectorsin universiti teknologi malaysia
publisher International Organization of Scientific Research (IOSR)
publishDate 2019
url http://eprints.utm.my/id/eprint/88558/1/NorehanMohdNor2019_TestperformanceofGammaSpectrometryCoAxial.pdf
http://eprints.utm.my/id/eprint/88558/
https://www.iosrjournals.org/iosr-jap/papers/Vol11-issue6/Series-1/I1106015866.pdf
_version_ 1687393588581761024