Part I: Development of sample preparation method for analysis of total iodine in air filters. Part II: Application of Inductively coupled plasma-mass spectrometry in analysis of radionuclides-uranium and plutonium

In Chapter one, an acid assisted closed vessel microwave digestion procedure was developed for the treatment of fine airborne particulate matter that allows elemental iodine to be solubilised simultaneously with metallic elements. This procedure is an improvement over existing methods as it allows t...