Direct comparison transfer of microwave power sensor calibration with an adaptor : modeling and evaluation

In this paper, calibration of a microwave power sensor with an adaptor is investigated with direct comparison transfer technique, and mathematically modeled using signal flow-graphs together with non-touching loop rules. The developed calibration model is then implemented practically with a 30 dB at...

全面介紹

Saved in:
書目詳細資料
Main Authors: Lin, Zhiping, Zhang, Qian., Meng, Yu Song., Shan, Yueyan.
其他作者: School of Electrical and Electronic Engineering
格式: Article
語言:English
出版: 2014
主題:
在線閱讀:https://hdl.handle.net/10356/84911
http://hdl.handle.net/10220/18439
http://www.jpier.org/PIERL/pier.php?paper=13012203
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:In this paper, calibration of a microwave power sensor with an adaptor is investigated with direct comparison transfer technique, and mathematically modeled using signal flow-graphs together with non-touching loop rules. The developed calibration model is then implemented practically with a 30 dB attenuator as the adaptor. Its performance is evaluated following the Guide to the Expression of Uncertainty in Measurement and also verified with the Monte Carlo method. Good agreements are observed with all the error |En| ≤ 0.25 over the whole frequency range (up to 18 GHz).