Preconcentration and Trace Analysis of Copper Based On Flow Injection Analysis Using Pyrocathecol Violet Modified XAD-16
Preconcentration method and flow injection analysis based copper trace analysis are based on the retention of Cu2+ ion in polystiren divinylbenzen (XAD-16) modified by Pyrocathecol Violet (PV), by using Flame Atomic Absorption Spectrometer (FAAS) as a detector. PV was retented well in pH 4 with the...
Saved in:
Main Author: | |
---|---|
Format: | Final Project |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/12208 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
Summary: | Preconcentration method and flow injection analysis based copper trace analysis are based on the retention of Cu2+ ion in polystiren divinylbenzen (XAD-16) modified by Pyrocathecol Violet (PV), by using Flame Atomic Absorption Spectrometer (FAAS) as a detector. PV was retented well in pH 4 with the retention capacity of 0,7896 mg PV / gram XAD-16. The analysis of Cu2+ ion retention in XAD-16 / PV in pH 5 was carried out by two method, the batch method and the column saturation level method. Analysis by the batch method resulted in the retention capacity of 0,6540 mg Cu2+ / gram XAD-16/PV, while the analysis by the column saturation level method resulted in the retention capacity of 0,0196 mg Cu2+ / gram XAD-16/PV. The minimum volume of 1M HNO3 for quantitative elution copper is 0,5 mL. Analytical performance of this developed method showed a good precision (%CV= 1,36) with limit detection = 7,66 ug L-1, sensitivity = 1,20 ug L-1), and sample recovery = 105,27 %. |
---|