Tools to be used to measure body weight and to measure gravity acceleration by using sensor LVDT is presented. By giving mass to load spot, ferro core (armature) at LVDT will be displace and generate the changing in voltage, This changing is used to convert voltage to weight. <br /> <br...
Saved in:
主要作者: | |
---|---|
格式: | Theses |
語言: | Indonesia |
在線閱讀: | https://digilib.itb.ac.id/gdl/view/20823 |
標簽: |
添加標簽
沒有標簽, 成為第一個標記此記錄!
|
總結: | Tools to be used to measure body weight and to measure gravity acceleration by using sensor LVDT is presented. By giving mass to load spot, ferro core (armature) at LVDT will be displace and generate the changing in voltage, This changing is used to convert voltage to weight. <br />
<br />
<br />
From the experiment, we get spring constant k=96,63 N/m with the sensivity of LVDT M=0,049 m/Volt. From linearity regression of the data experiment we get W = 4,5161 Vo + 4,1164. By comparing of weight measuring and counting we get error 7,545619 %, average of gravity acceleration is 9,32657 m/s2 with error 4,83082 %. |
---|