A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX
Athletes need a body system that is able to function properly and optimally to be able to produce the best performance. Many factors affect the function to run well and optimally, one important factor is good posture. Good posture makes the body symmetrical and balanced so that the distribution of l...
Saved in:
Main Author: | |
---|---|
Format: | Theses |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/24254 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
id |
id-itb.:24254 |
---|---|
spelling |
id-itb.:242542017-09-27T15:37:42ZA PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX SUKMANANDA SUPRAPTO (NIM : 23215310), SENA Indonesia Theses INSTITUT TEKNOLOGI BANDUNG https://digilib.itb.ac.id/gdl/view/24254 Athletes need a body system that is able to function properly and optimally to be able to produce the best performance. Many factors affect the function to run well and optimally, one important factor is good posture. Good posture makes the body symmetrical and balanced so that the distribution of loads spread evenly <br /> <br /> <br /> throughout the body. The current technological development of posture for foot pressure measurement early diagnostics is pedobarograph. <br /> <br /> <br /> In this research, the design of the footprint scanning system using pressure resistive sensor matrix (velostat) is used to determine the characteristics of <br /> <br /> <br /> resistive (velostat) sensors in the implementation and evaluation on the appropriate system for measuring the foot pressure distribution. <br /> <br /> <br /> The system design stage uses a diode as a current rectifier, where the phenomenon in which the output value of the sensor to the surrounding sensor in the matrix arrangement (cross talk) can be overcome. <br /> <br /> <br /> The system calibration stage uses a dynamometer of 1 - 15N range and there is a linear relationship between the compressive force and the sensor matrix conductance. There is a variation in the calibration value of each sensor on the sensor matrix where the value has no correlation to the coordinate of the sensor in the matrix which can be overcome by using the median filter. <br /> <br /> <br /> The system evaluation stage was obtained by comparing the sensor matrix with several other methods implemented in RSON and obtained the performance of the <br /> <br /> <br /> sensor matrix in the form of the comparison of the footprint area with the error value of 7.3% to 68.8%, the average pressure of 7.3% to 68.8%, the maximum <br /> <br /> <br /> pressure 37.6% to 70.7%, body weight of 0.4% to 12.1%, and arch-index with a specificity value of 0.3 and a sensitivity of 0.5. text |
institution |
Institut Teknologi Bandung |
building |
Institut Teknologi Bandung Library |
continent |
Asia |
country |
Indonesia Indonesia |
content_provider |
Institut Teknologi Bandung |
collection |
Digital ITB |
language |
Indonesia |
description |
Athletes need a body system that is able to function properly and optimally to be able to produce the best performance. Many factors affect the function to run well and optimally, one important factor is good posture. Good posture makes the body symmetrical and balanced so that the distribution of loads spread evenly <br />
<br />
<br />
throughout the body. The current technological development of posture for foot pressure measurement early diagnostics is pedobarograph. <br />
<br />
<br />
In this research, the design of the footprint scanning system using pressure resistive sensor matrix (velostat) is used to determine the characteristics of <br />
<br />
<br />
resistive (velostat) sensors in the implementation and evaluation on the appropriate system for measuring the foot pressure distribution. <br />
<br />
<br />
The system design stage uses a diode as a current rectifier, where the phenomenon in which the output value of the sensor to the surrounding sensor in the matrix arrangement (cross talk) can be overcome. <br />
<br />
<br />
The system calibration stage uses a dynamometer of 1 - 15N range and there is a linear relationship between the compressive force and the sensor matrix conductance. There is a variation in the calibration value of each sensor on the sensor matrix where the value has no correlation to the coordinate of the sensor in the matrix which can be overcome by using the median filter. <br />
<br />
<br />
The system evaluation stage was obtained by comparing the sensor matrix with several other methods implemented in RSON and obtained the performance of the <br />
<br />
<br />
sensor matrix in the form of the comparison of the footprint area with the error value of 7.3% to 68.8%, the average pressure of 7.3% to 68.8%, the maximum <br />
<br />
<br />
pressure 37.6% to 70.7%, body weight of 0.4% to 12.1%, and arch-index with a specificity value of 0.3 and a sensitivity of 0.5. |
format |
Theses |
author |
SUKMANANDA SUPRAPTO (NIM : 23215310), SENA |
spellingShingle |
SUKMANANDA SUPRAPTO (NIM : 23215310), SENA A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
author_facet |
SUKMANANDA SUPRAPTO (NIM : 23215310), SENA |
author_sort |
SUKMANANDA SUPRAPTO (NIM : 23215310), SENA |
title |
A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
title_short |
A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
title_full |
A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
title_fullStr |
A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
title_full_unstemmed |
A PROTOTYPE OF FOOT PRESSURE DISTRIBUTION SCANNER USING PRESSURE-RESISTIVE SENSOR MATRIX |
title_sort |
prototype of foot pressure distribution scanner using pressure-resistive sensor matrix |
url |
https://digilib.itb.ac.id/gdl/view/24254 |
_version_ |
1822921166411530240 |