Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage

Adityo Murdiarto has been done design and retrofit based thermoelectric exogenous insulin storage. The exogenous insulin storages was designed to store 12 @ 10 ml vial. The vials were placed inside aluminium shells and salt water was used as media in order to cool the shells. Temperature required...

Full description

Saved in:
Bibliographic Details
Main Author: Satia Graha, Bayu
Format: Final Project
Language:Indonesia
Subjects:
Online Access:https://digilib.itb.ac.id/gdl/view/39091
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Institut Teknologi Bandung
Language: Indonesia
id id-itb.:39091
spelling id-itb.:390912019-06-21T15:30:39ZPrototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage Satia Graha, Bayu Teknik (Rekayasa, enjinering dan kegiatan berkaitan) Indonesia Final Project INSTITUT TEKNOLOGI BANDUNG https://digilib.itb.ac.id/gdl/view/39091 Adityo Murdiarto has been done design and retrofit based thermoelectric exogenous insulin storage. The exogenous insulin storages was designed to store 12 @ 10 ml vial. The vials were placed inside aluminium shells and salt water was used as media in order to cool the shells. Temperature required to store insulin ranged between 2-8 oC. This exogenous insulin storages was designed to be able to maintain temperatures within the range and integrated with electricity system inside. The exogenous insulin storage box then made as a step to realize the designed. In the manufacturing process, there are some modifications made because of problems encountered during the process as well as the availability of goods on the market. Tests were conducted on the exogenous insulin storages to find out wether the temperature can be achieved. There are two types of testing, ie, cooler fluid temperatur and milk as representated of exogenous insulin using. Milk is used instead of insulin because it has a value that can represent the heat capacity of insulin. Milk temperature in the vial can reach 8°C in 190 minutes. From the results it was concluded that the design and manufacture can be used to store exogenous insulin. The result of this final project are expected to provide an overview of the process of making and testing of thermoelectric cooling box used to store exogenous insulin. text
institution Institut Teknologi Bandung
building Institut Teknologi Bandung Library
continent Asia
country Indonesia
Indonesia
content_provider Institut Teknologi Bandung
collection Digital ITB
language Indonesia
topic Teknik (Rekayasa, enjinering dan kegiatan berkaitan)
spellingShingle Teknik (Rekayasa, enjinering dan kegiatan berkaitan)
Satia Graha, Bayu
Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
description Adityo Murdiarto has been done design and retrofit based thermoelectric exogenous insulin storage. The exogenous insulin storages was designed to store 12 @ 10 ml vial. The vials were placed inside aluminium shells and salt water was used as media in order to cool the shells. Temperature required to store insulin ranged between 2-8 oC. This exogenous insulin storages was designed to be able to maintain temperatures within the range and integrated with electricity system inside. The exogenous insulin storage box then made as a step to realize the designed. In the manufacturing process, there are some modifications made because of problems encountered during the process as well as the availability of goods on the market. Tests were conducted on the exogenous insulin storages to find out wether the temperature can be achieved. There are two types of testing, ie, cooler fluid temperatur and milk as representated of exogenous insulin using. Milk is used instead of insulin because it has a value that can represent the heat capacity of insulin. Milk temperature in the vial can reach 8°C in 190 minutes. From the results it was concluded that the design and manufacture can be used to store exogenous insulin. The result of this final project are expected to provide an overview of the process of making and testing of thermoelectric cooling box used to store exogenous insulin.
format Final Project
author Satia Graha, Bayu
author_facet Satia Graha, Bayu
author_sort Satia Graha, Bayu
title Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
title_short Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
title_full Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
title_fullStr Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
title_full_unstemmed Prototyping and Testing the Redesigned and Retrofited Thermoelectric Based Exogenous Insulin Storage
title_sort prototyping and testing the redesigned and retrofited thermoelectric based exogenous insulin storage
url https://digilib.itb.ac.id/gdl/view/39091
_version_ 1822925195616190464