PENGEMBANGAN METODE SPEKTROFOTOMETRI UV-SINAR TAMPAK DENGAN PEREAKSI ALIZARIN MERAH S UNTUK PENENTUAN KADAR ALUMINIUM DAN MAGNESIUM DALAM SEDIAAN TABLET
Aluminium hydroxide and magnesium hydroxide are antacid drugs that increase pH stomach due to excessive production of stomach acid. These drug used in tablet to eliminate side effect from each other. The aim of this research was to develop a method for determination of aluminium and magnesium in...
Saved in:
Main Author: | |
---|---|
Format: | Final Project |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/44336 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
Summary: | Aluminium hydroxide and magnesium hydroxide are antacid drugs that increase pH stomach due
to excessive production of stomach acid. These drug used in tablet to eliminate side effect from
each other. The aim of this research was to develop a method for determination of aluminium
and magnesium in tablet using uv-visible spectrophotometry. The determination of aluminium
hydroxide and magnesium hydroxide is based on complex forming reactions between Al
3+
ion and
Mg
2+
with alizarin red S reagent in a certain pH. The research start by doing preliminary test,
reaction optimization, validation and application method. The result showed the composition of
reagents for aluminium determination were AMS 9 x 10
-3
% w/v, polyvinylpirolidon K25 1% w/v
and acetat buffer pH 4.5 1 M, while reagents composition for determination of magnesium was
not determined. The result of method validation shown that the method was linear in over the
range of aluminium 1.0 –7.0 µg/mL with correlation coefficient (r) 0.999; the detection limit of
the method was 0.2335 µg/mL while the quantization limit was 0.7783 µg/mL. Accuracy of the
method recovery was 100.55% ± 0.7335, while intraday and interday precision were shown by
relative standard deviation (RSD) < 2%. Based on result of the research, the developed method
has validated and can be applied to determine aluminium in tablet and can not be applied to
magnesium.
|
---|