DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS
In 2013, WHO reported that cardiovascular disease was a major cause of global mortality, contributing to 1/3 of its total, where 45% of said cases are caused by hypertension. One of the causes of hypertension is aldosterone-producing adenoma (APA), identified by the presence of tumor in the adren...
Saved in:
Main Author: | |
---|---|
Format: | Final Project |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/49882 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
id |
id-itb.:49882 |
---|---|
spelling |
id-itb.:498822020-09-21T11:16:14ZDESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS Muliany, Johanna Indonesia Final Project S-AVS, split-tip microcatheter, vein constriction, CFD, Kriging INSTITUT TEKNOLOGI BANDUNG https://digilib.itb.ac.id/gdl/view/49882 In 2013, WHO reported that cardiovascular disease was a major cause of global mortality, contributing to 1/3 of its total, where 45% of said cases are caused by hypertension. One of the causes of hypertension is aldosterone-producing adenoma (APA), identified by the presence of tumor in the adrenal glands. Therefore, a proper diagnosis of APA is necessary so that the patient can receive the appropriate medical treatment. Currently, the gold standard in the detection of APA in the body is the adrenal venous sampling (AVS) procedure. The most recent development of the AVS procedure is the S-AVS procedure, in which the blood sample is taken at the first branch of the central adrenal vein. Practically, there are several issues in the sampling process that are generally thought to be caused by venous constriction. To overcome this, the latest development in S-AVS is the use of a split-tip microcatheter, namely a microcatheter with a notch at one end. In practice, the microcatheter is considered capable of increasing the success of the sampling process. There are several studies which report that the use of a split-tip microcatheter reduces venous constriction. However, there is a need for a physical explanation of the effect of the use and design of the split-tip microcatheter in the S-AVS procedure. In this final project, an exploration of the split-tip microcatheter design is performed as an effort to explain its effects in the S-AVS procedure. Exploration of the microcatheter’s design variables, namely the amount and depth of the notches, was carried out in silico through computational fluid dynamics (CFD) simulations and metamodel construction of the system response. In the CFD simulation, model validation is carried out through the mesh independence test. The response metamodel construction was carried out using the Kriging regression method, so that the response of venous constriction with regards to the number and depth of notches could be described holistically. text |
institution |
Institut Teknologi Bandung |
building |
Institut Teknologi Bandung Library |
continent |
Asia |
country |
Indonesia Indonesia |
content_provider |
Institut Teknologi Bandung |
collection |
Digital ITB |
language |
Indonesia |
description |
In 2013, WHO reported that cardiovascular disease was a major cause of global
mortality, contributing to 1/3 of its total, where 45% of said cases are caused by
hypertension. One of the causes of hypertension is aldosterone-producing adenoma
(APA), identified by the presence of tumor in the adrenal glands. Therefore, a
proper diagnosis of APA is necessary so that the patient can receive the appropriate
medical treatment.
Currently, the gold standard in the detection of APA in the body is the adrenal
venous sampling (AVS) procedure. The most recent development of the AVS
procedure is the S-AVS procedure, in which the blood sample is taken at the first
branch of the central adrenal vein. Practically, there are several issues in the
sampling process that are generally thought to be caused by venous constriction. To
overcome this, the latest development in S-AVS is the use of a split-tip
microcatheter, namely a microcatheter with a notch at one end. In practice, the
microcatheter is considered capable of increasing the success of the sampling
process. There are several studies which report that the use of a split-tip
microcatheter reduces venous constriction. However, there is a need for a physical
explanation of the effect of the use and design of the split-tip microcatheter in the
S-AVS procedure.
In this final project, an exploration of the split-tip microcatheter design is performed
as an effort to explain its effects in the S-AVS procedure. Exploration of the
microcatheter’s design variables, namely the amount and depth of the notches, was
carried out in silico through computational fluid dynamics (CFD) simulations and
metamodel construction of the system response. In the CFD simulation, model
validation is carried out through the mesh independence test. The response
metamodel construction was carried out using the Kriging regression method, so
that the response of venous constriction with regards to the number and depth of
notches could be described holistically. |
format |
Final Project |
author |
Muliany, Johanna |
spellingShingle |
Muliany, Johanna DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
author_facet |
Muliany, Johanna |
author_sort |
Muliany, Johanna |
title |
DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
title_short |
DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
title_full |
DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
title_fullStr |
DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
title_full_unstemmed |
DESIGN EXPLORATION OF MICROCATHETER TIP GEOMETRY FOR SEGMENTAL ADRENAL VENOUS SAMPLING USING COMPUTATIONAL FLUID DYNAMICS |
title_sort |
design exploration of microcatheter tip geometry for segmental adrenal venous sampling using computational fluid dynamics |
url |
https://digilib.itb.ac.id/gdl/view/49882 |
_version_ |
1822272173307330560 |