ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF AMLODIPINE BESYLATE AND ITS IMPURITIES IN TABLETS DOSAGE FORM
Amlodipine besylate belongs to the class of oral dihydropyridine calcium channel blockers that can be used to hypertension and angina therapy. Amlodipine as a pharmaceutical preparation must meet the criteria for a good pharmaceutical preparation, namely safe, efficacious and quality. In terms of...
Saved in:
Main Author: | |
---|---|
Format: | Theses |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/61934 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
Summary: | Amlodipine besylate belongs to the class of oral dihydropyridine calcium channel
blockers that can be used to hypertension and angina therapy. Amlodipine as a
pharmaceutical preparation must meet the criteria for a good pharmaceutical
preparation, namely safe, efficacious and quality. In terms of safety, the presence
of organic impurity in pharmaceutical products can affect the safety of medicinal
products. Because the difficultness to obtain a standard of impurity, it is
necessary to develop an analytical method to analyze the presence of
contamination by stress testing. The analytical method was developed using a
High Performance Liquid Chromatography (HPLC) instrument with Eclipse Plus
C18 column 5 m (150 x 4.6 mm), mobile phase triethylamine (TEA): methanol in a
ratio of 40:60 with isocratic elution and a flow rate of 1 mL/minutes and a
wavelength of 237 nm with an injection volume of 10 µL. The results of the
method validation showed good linearity with R
2
of 0.9996, detection limit of
18.46 µg/mL and quantization limit value of 61.54 µg/mL. The value of the
coefficient of variation on the precision parameter is 1.18-1.26%. The percent
recovery accuracy value is 93.07-105.44% and the range value for proportional
samples is in the concentration range of 24-84 µg/ml with R
2
= 0.9993. The
HPLC system meets all requirements for acceptance of the system conformity test.
The results of this experiment showed that there were 2 out of 5 peaks of
degradation products whose chemical structure was known. The application of
the analytical method on commercial samples showed levels between 90.79-
95.68% and there was no impurity in the product samples.
|
---|