Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose

Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepare...

Full description

Saved in:
Bibliographic Details
Main Authors: Miratul Khasanah, .-, Alfa Akustia Widati, .-, Usreg Sri Handajani, .-, Masfah Raudlotus Shofiyyah, .-, Sabrina Aulia Rakhma, .-, Herwin Predianto, .-
Format: Book Section PeerReviewed
Language:English
English
English
English
Published: AIP Publishing 2019
Subjects:
Online Access:https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf
https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf
https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf
https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf
https://repository.unair.ac.id/125998/
https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universitas Airlangga
Language: English
English
English
English
id id-langga.125998
record_format dspace
spelling id-langga.1259982023-06-05T01:12:26Z https://repository.unair.ac.id/125998/ Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose Miratul Khasanah, .- Alfa Akustia Widati, .- Usreg Sri Handajani, .- Masfah Raudlotus Shofiyyah, .- Sabrina Aulia Rakhma, .- Herwin Predianto, .- Q Science Q Science (General) QD Chemistry QD71-142 Analytical chemistry Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepared by mixing activated carbon, imprinted zeolite, and paraffin granule with mass ratio 9:4:7 from the measurement in range of 10−5 – 10−2 M is with Nernst factor of 28.6 mV/decade, and the detection limit of 4.79 x 10-5 M. The electrode was able to respond the glucose molecules in solution quickly (<30 s), stable for more than 9 weeks (130 times usage) and selective to glucose molecules. The urea, uric acid, and creatine with various concentrations did not interfere to glucose analysis by potentiometry using the electrode. Applying of the electrode for glucose analysis in spiked blood serum samples showed that it had recovery of 91.73±3.48% (n=5), while its accuracy against spectrophotometry method as commonly used method in medical field was 90.58±4.65% (n=5). Based on its performance, potentiometry using the imprinted zeolite modified carbon paste electrode is recommended as an alternative method for routine blood glucose analysis in the medical field. AIP Publishing 2019 Book Section PeerReviewed text en https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf text en https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf text en https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf text en https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf Miratul Khasanah, .- and Alfa Akustia Widati, .- and Usreg Sri Handajani, .- and Masfah Raudlotus Shofiyyah, .- and Sabrina Aulia Rakhma, .- and Herwin Predianto, .- (2019) Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose. In: The 14th Joint Conference on Chemistry 2019. AIP Publishing, Surakarta, Indonesia, 020011-1. ISBN ISSN:0094-243X E-ISSN:1551-7616, https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext
institution Universitas Airlangga
building Universitas Airlangga Library
continent Asia
country Indonesia
Indonesia
content_provider Universitas Airlangga Library
collection UNAIR Repository
language English
English
English
English
topic Q Science
Q Science (General)
QD Chemistry
QD71-142 Analytical chemistry
spellingShingle Q Science
Q Science (General)
QD Chemistry
QD71-142 Analytical chemistry
Miratul Khasanah, .-
Alfa Akustia Widati, .-
Usreg Sri Handajani, .-
Masfah Raudlotus Shofiyyah, .-
Sabrina Aulia Rakhma, .-
Herwin Predianto, .-
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
description Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepared by mixing activated carbon, imprinted zeolite, and paraffin granule with mass ratio 9:4:7 from the measurement in range of 10−5 – 10−2 M is with Nernst factor of 28.6 mV/decade, and the detection limit of 4.79 x 10-5 M. The electrode was able to respond the glucose molecules in solution quickly (<30 s), stable for more than 9 weeks (130 times usage) and selective to glucose molecules. The urea, uric acid, and creatine with various concentrations did not interfere to glucose analysis by potentiometry using the electrode. Applying of the electrode for glucose analysis in spiked blood serum samples showed that it had recovery of 91.73±3.48% (n=5), while its accuracy against spectrophotometry method as commonly used method in medical field was 90.58±4.65% (n=5). Based on its performance, potentiometry using the imprinted zeolite modified carbon paste electrode is recommended as an alternative method for routine blood glucose analysis in the medical field.
format Book Section
PeerReviewed
author Miratul Khasanah, .-
Alfa Akustia Widati, .-
Usreg Sri Handajani, .-
Masfah Raudlotus Shofiyyah, .-
Sabrina Aulia Rakhma, .-
Herwin Predianto, .-
author_facet Miratul Khasanah, .-
Alfa Akustia Widati, .-
Usreg Sri Handajani, .-
Masfah Raudlotus Shofiyyah, .-
Sabrina Aulia Rakhma, .-
Herwin Predianto, .-
author_sort Miratul Khasanah, .-
title Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
title_short Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
title_full Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
title_fullStr Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
title_full_unstemmed Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
title_sort imprinted zeolite modified carbon paste electrode as a selective potentiometric sensor for blood glucose
publisher AIP Publishing
publishDate 2019
url https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf
https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf
https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf
https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf
https://repository.unair.ac.id/125998/
https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext
_version_ 1768017631096340480