Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepare...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Book Section PeerReviewed |
Language: | English English English English |
Published: |
AIP Publishing
2019
|
Subjects: | |
Online Access: | https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf https://repository.unair.ac.id/125998/ https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universitas Airlangga |
Language: | English English English English |
id |
id-langga.125998 |
---|---|
record_format |
dspace |
spelling |
id-langga.1259982023-06-05T01:12:26Z https://repository.unair.ac.id/125998/ Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose Miratul Khasanah, .- Alfa Akustia Widati, .- Usreg Sri Handajani, .- Masfah Raudlotus Shofiyyah, .- Sabrina Aulia Rakhma, .- Herwin Predianto, .- Q Science Q Science (General) QD Chemistry QD71-142 Analytical chemistry Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepared by mixing activated carbon, imprinted zeolite, and paraffin granule with mass ratio 9:4:7 from the measurement in range of 10−5 – 10−2 M is with Nernst factor of 28.6 mV/decade, and the detection limit of 4.79 x 10-5 M. The electrode was able to respond the glucose molecules in solution quickly (<30 s), stable for more than 9 weeks (130 times usage) and selective to glucose molecules. The urea, uric acid, and creatine with various concentrations did not interfere to glucose analysis by potentiometry using the electrode. Applying of the electrode for glucose analysis in spiked blood serum samples showed that it had recovery of 91.73±3.48% (n=5), while its accuracy against spectrophotometry method as commonly used method in medical field was 90.58±4.65% (n=5). Based on its performance, potentiometry using the imprinted zeolite modified carbon paste electrode is recommended as an alternative method for routine blood glucose analysis in the medical field. AIP Publishing 2019 Book Section PeerReviewed text en https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf text en https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf text en https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf text en https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf Miratul Khasanah, .- and Alfa Akustia Widati, .- and Usreg Sri Handajani, .- and Masfah Raudlotus Shofiyyah, .- and Sabrina Aulia Rakhma, .- and Herwin Predianto, .- (2019) Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose. In: The 14th Joint Conference on Chemistry 2019. AIP Publishing, Surakarta, Indonesia, 020011-1. ISBN ISSN:0094-243X E-ISSN:1551-7616, https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext |
institution |
Universitas Airlangga |
building |
Universitas Airlangga Library |
continent |
Asia |
country |
Indonesia Indonesia |
content_provider |
Universitas Airlangga Library |
collection |
UNAIR Repository |
language |
English English English English |
topic |
Q Science Q Science (General) QD Chemistry QD71-142 Analytical chemistry |
spellingShingle |
Q Science Q Science (General) QD Chemistry QD71-142 Analytical chemistry Miratul Khasanah, .- Alfa Akustia Widati, .- Usreg Sri Handajani, .- Masfah Raudlotus Shofiyyah, .- Sabrina Aulia Rakhma, .- Herwin Predianto, .- Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
description |
Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepared by mixing activated carbon, imprinted zeolite, and paraffin granule with mass ratio 9:4:7 from the measurement in range of 10−5 – 10−2 M is with Nernst factor of 28.6 mV/decade, and the detection limit of 4.79 x 10-5 M. The electrode was able to respond the glucose molecules in solution quickly (<30 s), stable for more than 9 weeks (130 times usage) and selective to glucose molecules. The urea, uric acid, and creatine with various concentrations did not interfere to glucose analysis by potentiometry using the electrode. Applying of the electrode for glucose analysis in spiked blood serum samples showed that it had recovery of 91.73±3.48% (n=5), while its accuracy against spectrophotometry method as commonly used method in medical field was 90.58±4.65% (n=5). Based on its performance, potentiometry using the imprinted zeolite modified carbon paste electrode is recommended as an alternative method for routine blood glucose analysis in the medical field. |
format |
Book Section PeerReviewed |
author |
Miratul Khasanah, .- Alfa Akustia Widati, .- Usreg Sri Handajani, .- Masfah Raudlotus Shofiyyah, .- Sabrina Aulia Rakhma, .- Herwin Predianto, .- |
author_facet |
Miratul Khasanah, .- Alfa Akustia Widati, .- Usreg Sri Handajani, .- Masfah Raudlotus Shofiyyah, .- Sabrina Aulia Rakhma, .- Herwin Predianto, .- |
author_sort |
Miratul Khasanah, .- |
title |
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
title_short |
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
title_full |
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
title_fullStr |
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
title_full_unstemmed |
Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose |
title_sort |
imprinted zeolite modified carbon paste electrode as a selective potentiometric sensor for blood glucose |
publisher |
AIP Publishing |
publishDate |
2019 |
url |
https://repository.unair.ac.id/125998/1/C27.%20Fulltext.pdf https://repository.unair.ac.id/125998/2/C27.%20Penilaian%20dan%20Validasi.pdf https://repository.unair.ac.id/125998/3/C27.%20Similarity.pdf https://repository.unair.ac.id/125998/7/C27.%20Bukti%20Korespondensi_rev.pdf https://repository.unair.ac.id/125998/ https://pubs.aip.org/aip/acp/article-abstract/2237/1/020011/1007408/Imprinted-zeolite-modified-carbon-paste-electrode?redirectedFrom=fulltext |
_version_ |
1768017631096340480 |