PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY
Comparative on image quality has been done for digital fluoroscopic radiography a frame grabber (DFR) and by a computed radiography (CR) and its relation to the entrace skin exposure (ESE) value. ESE measurement was using free in air method by a personal alarm dosimeter. The phantom object of antrom...
Saved in:
Main Authors: | , |
---|---|
Format: | Theses and Dissertations NonPeerReviewed |
Published: |
[Yogyakarta] : Universitas Gadjah Mada
2011
|
Subjects: | |
Online Access: | https://repository.ugm.ac.id/97388/ http://etd.ugm.ac.id/index.php?mod=penelitian_detail&sub=PenelitianDetail&act=view&typ=html&buku_id=53651 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universitas Gadjah Mada |
id |
id-ugm-repo.97388 |
---|---|
record_format |
dspace |
spelling |
id-ugm-repo.973882016-03-04T08:50:24Z https://repository.ugm.ac.id/97388/ PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY , SITI DARYATI , Dr. Gede Bayu Suparta ETD Comparative on image quality has been done for digital fluoroscopic radiography a frame grabber (DFR) and by a computed radiography (CR) and its relation to the entrace skin exposure (ESE) value. ESE measurement was using free in air method by a personal alarm dosimeter. The phantom object of antromorphic phantom pelvis was used. Digitization process was using a frame grabber that was connected to the output of a fluoroscopy imaging system. Results from digital fluoroscopy using DrGrabber was compared to the results from CR. Contrast analysis based on histogram and line profile was using ImageJ software. The ESE value for CR and digitized by the frame grabber increased as the voltage increased. The ESE value of CR lowest 54.35 ± 2 and highest 111.36 ± 2 while The ESE value of DFR lowest 17.48 ± 1 and highest 117.45 ± 2. The ESE value of DFR (quartil 1 = 30,60 mR, quartil 2 = 49,60 mR and quartil 3 = 78,01 mR)was lower than the ESE value of CR (quartil 1 = 61,10 mR, quartil 2 = 74,69 mR and quartil 3 = 90,55 mR). These results are lower than the quartil 1, 2, dan 3 of ESE value for abdomen AP diagnostic of AAPM Report No.31 (2005). The voltage increase in CR provided no contrast difference based on color gradation, while a better contrast difference are obtained by DFR. [Yogyakarta] : Universitas Gadjah Mada 2011 Thesis NonPeerReviewed , SITI DARYATI and , Dr. Gede Bayu Suparta (2011) PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY. UNSPECIFIED thesis, UNSPECIFIED. http://etd.ugm.ac.id/index.php?mod=penelitian_detail&sub=PenelitianDetail&act=view&typ=html&buku_id=53651 |
institution |
Universitas Gadjah Mada |
building |
UGM Library |
country |
Indonesia |
collection |
Repository Civitas UGM |
topic |
ETD |
spellingShingle |
ETD , SITI DARYATI , Dr. Gede Bayu Suparta PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
description |
Comparative on image quality has been done for digital fluoroscopic radiography
a frame grabber (DFR) and by a computed radiography (CR) and its relation to the
entrace skin exposure (ESE) value. ESE measurement was using free in air method
by a personal alarm dosimeter. The phantom object of antromorphic phantom pelvis
was used. Digitization process was using a frame grabber that was connected to the
output of a fluoroscopy imaging system. Results from digital fluoroscopy using
DrGrabber was compared to the results from CR. Contrast analysis based on
histogram and line profile was using ImageJ software.
The ESE value for CR and digitized by the frame grabber increased as the
voltage increased. The ESE value of CR lowest 54.35 ± 2 and highest 111.36 ± 2
while The ESE value of DFR lowest 17.48 ± 1 and highest 117.45 ± 2. The ESE
value of DFR (quartil 1 = 30,60 mR, quartil 2 = 49,60 mR and quartil 3 = 78,01
mR)was lower than the ESE value of CR (quartil 1 = 61,10 mR, quartil 2 = 74,69 mR
and quartil 3 = 90,55 mR). These results are lower than the quartil 1, 2, dan 3 of ESE
value for abdomen AP diagnostic of AAPM Report No.31 (2005). The voltage
increase in CR provided no contrast difference based on color gradation, while a
better contrast difference are obtained by DFR. |
format |
Theses and Dissertations NonPeerReviewed |
author |
, SITI DARYATI , Dr. Gede Bayu Suparta |
author_facet |
, SITI DARYATI , Dr. Gede Bayu Suparta |
author_sort |
, SITI DARYATI |
title |
PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA
HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
title_short |
PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA
HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
title_full |
PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA
HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
title_fullStr |
PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA
HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
title_full_unstemmed |
PERBANDINGAN KUALITAS CITRA RADIOGRAFI DIGITAL ANTARA
HASIL DIGITISASI FRAME GRABBER DAN COMPUTED RADIOGRAPHY |
title_sort |
perbandingan kualitas citra radiografi digital antara
hasil digitisasi frame grabber dan computed radiography |
publisher |
[Yogyakarta] : Universitas Gadjah Mada |
publishDate |
2011 |
url |
https://repository.ugm.ac.id/97388/ http://etd.ugm.ac.id/index.php?mod=penelitian_detail&sub=PenelitianDetail&act=view&typ=html&buku_id=53651 |
_version_ |
1681230164201046016 |