Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants
Background: Strength testing is one of the primary assessment tools for physical strength examination, and an increase in strength is the aim of many rehabilitation programs. Manual muscle testing (MMT) is the most commonly used strength testing method in clinical settings. It is a subjective measur...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
INTI International University
2023
|
Subjects: | |
Online Access: | http://eprints.intimal.edu.my/1788/1/ij2023_39.pdf http://eprints.intimal.edu.my/1788/ https://intijournal.intimal.edu.my |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | INTI International University |
Language: | English |
id |
my-inti-eprints.1788 |
---|---|
record_format |
eprints |
spelling |
my-inti-eprints.17882023-09-13T09:50:59Z http://eprints.intimal.edu.my/1788/ Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants S.J., Kanimozhiselvi M., Sindhuja J., Radhika Q Science (General) QP Physiology Background: Strength testing is one of the primary assessment tools for physical strength examination, and an increase in strength is the aim of many rehabilitation programs. Manual muscle testing (MMT) is the most commonly used strength testing method in clinical settings. It is a subjective measurement technique whereby the tester applies resistance to a maximum voluntary muscle contraction. The reliability of the Fitmust hand-held dynamometer is not previously reported in the literature. Aims and Objectives: The study aims to test the reliability of the Fitmust hand-held dynamometer for lower limb muscles in comparison with the Jamar Hand Held dynamometer. Methods: Maximum voluntary isometric contraction of hip extension, knee extension, and plantar flexion is measured by two different therapists on different days using fit must hand-held dynamometer for 40 healthy participants of age category 18- 30 years. Result: Reliability was evaluated by computing Intra-class Correlation Coefficients (ICC), which analyse the consistency between two or more quantitative measures, and Cronbach's Alpha which measured agreement. Conclusion: Inter-tester reliability data obtained for Assessor 1 and 2 were consistent for knee measurements. However, data for the hip and ankle varied considerably. Assessors 1 and 2 demonstrated a 'moderate to excellent' association for knee Measurements. INTI International University 2023-09 Article PeerReviewed text en cc_by_4 http://eprints.intimal.edu.my/1788/1/ij2023_39.pdf S.J., Kanimozhiselvi and M., Sindhuja and J., Radhika (2023) Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants. INTI JOURNAL, 2023 (39). pp. 1-7. ISSN e2600-7320 https://intijournal.intimal.edu.my |
institution |
INTI International University |
building |
INTI Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
INTI International University |
content_source |
INTI Institutional Repository |
url_provider |
http://eprints.intimal.edu.my |
language |
English |
topic |
Q Science (General) QP Physiology |
spellingShingle |
Q Science (General) QP Physiology S.J., Kanimozhiselvi M., Sindhuja J., Radhika Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
description |
Background: Strength testing is one of the primary assessment tools for physical strength examination, and an increase in strength is the aim of many rehabilitation programs. Manual muscle testing (MMT) is the most commonly used strength testing method in clinical settings. It is a subjective measurement technique whereby the tester applies resistance to a maximum voluntary muscle contraction. The reliability of the Fitmust hand-held dynamometer is not previously reported in the literature.
Aims and Objectives: The study aims to test the reliability of the Fitmust hand-held dynamometer for lower limb muscles in comparison with the Jamar Hand Held dynamometer.
Methods: Maximum voluntary isometric contraction of hip extension, knee extension, and plantar flexion is measured by two different therapists on different days using fit must hand-held dynamometer for 40 healthy participants of age category 18- 30 years.
Result: Reliability was evaluated by computing Intra-class Correlation Coefficients (ICC), which analyse the consistency between two or more quantitative measures, and Cronbach's Alpha which measured agreement.
Conclusion: Inter-tester reliability data obtained for Assessor 1 and 2 were consistent for knee measurements. However, data for the hip and ankle varied considerably. Assessors 1 and 2 demonstrated a 'moderate to excellent' association for knee Measurements. |
format |
Article |
author |
S.J., Kanimozhiselvi M., Sindhuja J., Radhika |
author_facet |
S.J., Kanimozhiselvi M., Sindhuja J., Radhika |
author_sort |
S.J., Kanimozhiselvi |
title |
Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
title_short |
Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
title_full |
Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
title_fullStr |
Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
title_full_unstemmed |
Inter Tester Reliability of Fitmust Hand Held Dynamometer among Healthy Participants |
title_sort |
inter tester reliability of fitmust hand held dynamometer among healthy participants |
publisher |
INTI International University |
publishDate |
2023 |
url |
http://eprints.intimal.edu.my/1788/1/ij2023_39.pdf http://eprints.intimal.edu.my/1788/ https://intijournal.intimal.edu.my |
_version_ |
1778167414299557888 |