Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital
In this study, 6 MV photon beam of TrueBeam STx Varian LINAC with Flattening Filter Free (FFF) was simulated using PRIMO code. The depth dose profiles for various jaws open fields and cross beam profiles for various depths inside water phantom were determined using Monte Carlo (MC) simulation techni...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Penerbit Universiti Kebangsaan Malaysia
2017
|
Online Access: | http://journalarticle.ukm.my/11365/1/08%20M.%20Arif%20Efendi.pdf http://journalarticle.ukm.my/11365/ http://www.ukm.my/jsm/english_journals/vol46num9_2017/contentsVol46num9_2017.html |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Kebangsaan Malaysia |
Language: | English |
id |
my-ukm.journal.11365 |
---|---|
record_format |
eprints |
spelling |
my-ukm.journal.113652018-02-12T03:03:23Z http://journalarticle.ukm.my/11365/ Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital M. Arif Efendi, Amporn Funsian, Thawat Chittrakarn, Tripob Bhongsuwan, In this study, 6 MV photon beam of TrueBeam STx Varian LINAC with Flattening Filter Free (FFF) was simulated using PRIMO code. The depth dose profiles for various jaws open fields and cross beam profiles for various depths inside water phantom were determined using Monte Carlo (MC) simulation technique and validated with experimental result. The experiments were performed using the Source to Surface Distance (SSD) technique with a 100 cm distance from target to the surface of water. Simulation used 109 histories with the same configurations with experiments. The depth dose profiles and cross beam profiles of 6 MV FFF photon beam was determined using MC simulations and compared with experimental results. The results showed that depth dose profiles and cross beam profiles by MC simulation accurately matched with experimental results. The best result of depth dose profile was obtained at 10×10 cm2 jaws open field with 98.53% passing criterion whereas cross beam profile was obtained at 10 cm depth inside water phantom with 88.96% passing criterion.The discrepancies were caused by scatter of particle and incompatibility of primary beam in PRIMO with experiment. Penerbit Universiti Kebangsaan Malaysia 2017-09 Article PeerReviewed application/pdf en http://journalarticle.ukm.my/11365/1/08%20M.%20Arif%20Efendi.pdf M. Arif Efendi, and Amporn Funsian, and Thawat Chittrakarn, and Tripob Bhongsuwan, (2017) Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital. Sains Malaysiana, 46 (9). pp. 1407-1411. ISSN 0126-6039 http://www.ukm.my/jsm/english_journals/vol46num9_2017/contentsVol46num9_2017.html |
institution |
Universiti Kebangsaan Malaysia |
building |
Perpustakaan Tun Sri Lanang Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Kebangsaan Malaysia |
content_source |
UKM Journal Article Repository |
url_provider |
http://journalarticle.ukm.my/ |
language |
English |
description |
In this study, 6 MV photon beam of TrueBeam STx Varian LINAC with Flattening Filter Free (FFF) was simulated using PRIMO code. The depth dose profiles for various jaws open fields and cross beam profiles for various depths inside water phantom were determined using Monte Carlo (MC) simulation technique and validated with experimental result. The experiments were performed using the Source to Surface Distance (SSD) technique with a 100 cm distance from target to the surface of water. Simulation used 109 histories with the same configurations with experiments. The depth dose profiles and cross beam profiles of 6 MV FFF photon beam was determined using MC simulations and compared with experimental results. The results showed that depth dose profiles and cross beam profiles by MC simulation accurately matched with experimental results. The best result of depth dose profile was obtained at 10×10 cm2 jaws open field with 98.53% passing criterion whereas cross beam profile was obtained at 10 cm depth inside water phantom with 88.96% passing criterion.The discrepancies were caused by scatter of particle and incompatibility of primary beam in PRIMO with experiment. |
format |
Article |
author |
M. Arif Efendi, Amporn Funsian, Thawat Chittrakarn, Tripob Bhongsuwan, |
spellingShingle |
M. Arif Efendi, Amporn Funsian, Thawat Chittrakarn, Tripob Bhongsuwan, Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
author_facet |
M. Arif Efendi, Amporn Funsian, Thawat Chittrakarn, Tripob Bhongsuwan, |
author_sort |
M. Arif Efendi, |
title |
Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
title_short |
Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
title_full |
Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
title_fullStr |
Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
title_full_unstemmed |
Monte Carlo simulation of 6 MV flattening filter free photon beam of TrueBeam STx LINAC at Songklanagarind Hospital |
title_sort |
monte carlo simulation of 6 mv flattening filter free photon beam of truebeam stx linac at songklanagarind hospital |
publisher |
Penerbit Universiti Kebangsaan Malaysia |
publishDate |
2017 |
url |
http://journalarticle.ukm.my/11365/1/08%20M.%20Arif%20Efendi.pdf http://journalarticle.ukm.my/11365/ http://www.ukm.my/jsm/english_journals/vol46num9_2017/contentsVol46num9_2017.html |
_version_ |
1643738467813818368 |