Study on different anode/filter combinations and tube voltage values on image quality in mammography
The number of women around the world diagnosed with breast cancer is increasing every year. Mammography is the best equipment for the early diagnosis of breast cancer. Imaging the breast is challenging since there are only small differences in X-ray attenuation in the various breast soft tissue st...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Penerbit Universiti Sultan Zainal Abidin
2020
|
Subjects: | |
Online Access: | http://eprints.unisza.edu.my/7472/1/FH02-FSK-20-48350.pdf http://eprints.unisza.edu.my/7472/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Sultan Zainal Abidin |
Language: | English |
id |
my-unisza-ir.7472 |
---|---|
record_format |
eprints |
spelling |
my-unisza-ir.74722022-06-08T01:43:39Z http://eprints.unisza.edu.my/7472/ Study on different anode/filter combinations and tube voltage values on image quality in mammography Fairuz, Mohd Nasir Nurul Syazwina, Mohamed Nur Nazifah, Mohd Hashim QC Physics R Medicine (General) The number of women around the world diagnosed with breast cancer is increasing every year. Mammography is the best equipment for the early diagnosis of breast cancer. Imaging the breast is challenging since there are only small differences in X-ray attenuation in the various breast soft tissue structures. Poor quality images reduce the detection rate of early breast cancer and the patient's chances of survival. Thus, this study aims to determine how the different anode/filter types and tube voltage values affect the image quality in mammography. Two different digital radiography systems of mammography were used. Firstly, using direct digital mammography (DDM) with Tungsten (W) as the anode, Rhodium (Rh) and Silver (Ag) as filters. Secondly, computed radiography (CR) mammogram system had been operated with Molybdenum (Mo) as both anode and filter to image the RMI 156 phantom replacing the human breast. The phantom was exposed using anode/filter combinations of Mo/Mo, W/Rh, and W/Ag, at the different tube kVp values of 24, 26, 28, 30 and 32 kV. The resultant image was evaluated by counting the number of test object visible in the image. The W/Rh anode/filter and low tube kVp values gave the best image quality in mammography as they increase the visibility of image details. Penerbit Universiti Sultan Zainal Abidin 2020-12 Article PeerReviewed text en http://eprints.unisza.edu.my/7472/1/FH02-FSK-20-48350.pdf Fairuz, Mohd Nasir and Nurul Syazwina, Mohamed and Nur Nazifah, Mohd Hashim (2020) Study on different anode/filter combinations and tube voltage values on image quality in mammography. Asian Journal of Medicine and Biomedicine, 4 (SI 1). pp. 74-78. ISSN 2600-8173 |
institution |
Universiti Sultan Zainal Abidin |
building |
UNISZA Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Sultan Zainal Abidin |
content_source |
UNISZA Institutional Repository |
url_provider |
https://eprints.unisza.edu.my/ |
language |
English |
topic |
QC Physics R Medicine (General) |
spellingShingle |
QC Physics R Medicine (General) Fairuz, Mohd Nasir Nurul Syazwina, Mohamed Nur Nazifah, Mohd Hashim Study on different anode/filter combinations and tube voltage values on image quality in mammography |
description |
The number of women around the world diagnosed with breast cancer is increasing every year. Mammography is
the best equipment for the early diagnosis of breast cancer. Imaging the breast is challenging since there are only
small differences in X-ray attenuation in the various breast soft tissue structures. Poor quality images reduce the
detection rate of early breast cancer and the patient's chances of survival. Thus, this study aims to determine how
the different anode/filter types and tube voltage values affect the image quality in mammography. Two different
digital radiography systems of mammography were used. Firstly, using direct digital mammography (DDM) with
Tungsten (W) as the anode, Rhodium (Rh) and Silver (Ag) as filters. Secondly, computed radiography (CR)
mammogram system had been operated with Molybdenum (Mo) as both anode and filter to image the RMI 156
phantom replacing the human breast. The phantom was exposed using anode/filter combinations of Mo/Mo,
W/Rh, and W/Ag, at the different tube kVp values of 24, 26, 28, 30 and 32 kV. The resultant image was evaluated
by counting the number of test object visible in the image. The W/Rh anode/filter and low tube kVp values gave
the best image quality in mammography as they increase the visibility of image details. |
format |
Article |
author |
Fairuz, Mohd Nasir Nurul Syazwina, Mohamed Nur Nazifah, Mohd Hashim |
author_facet |
Fairuz, Mohd Nasir Nurul Syazwina, Mohamed Nur Nazifah, Mohd Hashim |
author_sort |
Fairuz, Mohd Nasir |
title |
Study on different anode/filter combinations and tube voltage values on image quality in mammography |
title_short |
Study on different anode/filter combinations and tube voltage values on image quality in mammography |
title_full |
Study on different anode/filter combinations and tube voltage values on image quality in mammography |
title_fullStr |
Study on different anode/filter combinations and tube voltage values on image quality in mammography |
title_full_unstemmed |
Study on different anode/filter combinations and tube voltage values on image quality in mammography |
title_sort |
study on different anode/filter combinations and tube voltage values on image quality in mammography |
publisher |
Penerbit Universiti Sultan Zainal Abidin |
publishDate |
2020 |
url |
http://eprints.unisza.edu.my/7472/1/FH02-FSK-20-48350.pdf http://eprints.unisza.edu.my/7472/ |
_version_ |
1735391141088460800 |