Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)

In radiography, inconsistencies in patients' measured entrance skin dose (ESD) exist. There is no published research on the bucky table induced backscattered radiation dose (BTI-BSD). Thus, we aimed to ascertain ESD, calculate the BTI-BSD in abdominal radiography with a nanoDot OSLD, and compar...

Full description

Saved in:
Bibliographic Details
Main Authors: Sayed, Inayatullah Shah, Roslan, Nurul Shuhada, Syed, Waliullah Shah
Format: Article
Language:English
English
Published: Cureus, Inc 2023
Subjects:
Online Access:http://irep.iium.edu.my/103580/2/103580_Entrance%20skin%20dose%20%28ESD%29%20and%20bucky%20table.pdf
http://irep.iium.edu.my/103580/3/WOS%202023.pdf
http://irep.iium.edu.my/103580/
https://www.cureus.com/articles/131554-entrance-skin-dose-esd-and-bucky-table-induced-backscattered-dose-bti-bsd-in-abdominal-radiography-with-nanodot-optically-stimulated-luminescence-dosimeter-osld#!/authors
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Islam Antarabangsa Malaysia
Language: English
English
id my.iium.irep.103580
record_format dspace
spelling my.iium.irep.1035802023-05-03T04:50:32Z http://irep.iium.edu.my/103580/ Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD) Sayed, Inayatullah Shah Roslan, Nurul Shuhada Syed, Waliullah Shah QC Physics In radiography, inconsistencies in patients' measured entrance skin dose (ESD) exist. There is no published research on the bucky table induced backscattered radiation dose (BTI-BSD). Thus, we aimed to ascertain ESD, calculate the BTI-BSD in abdominal radiography with a nanoDot OSLD, and compare the ESD results with the published data. A Kyoto Kagaku PBU-50 phantom (Kyoto, Japan) in an antero-posterior supine position was exposed, selecting a protocol used for abdominal radiography. The central ray of x-ray beam was pointed at the surface of abdomen at the navel, where a nanoDot dosimeter was placed to measure ESD. For the BTI-BSD, exit dose (ED) was determined by placing a second dosimeter on the exact opposite side (backside) of the phantom from the dosimeter used to determine (ESD) with and without bucky table at identical exposure parameters. The BTI-BSD was calculated as the difference between ED with and without bucky table. The ESD, ED, and BTI-BSD were measured in milligray (mGy). ESD mean values with and without bucky table were 1.97 mGy and 1.84 mGy, whereas ED values were 0.062 mGy and 0.052 mGy, respectively. Results show 2-26% lower ESD values with nanoDot OSLD. The BTI-BSD mean value was found to be approximately 0.01 mGy. A local dose reference level (LDRL) can be established using ESD data to safeguard patients from unnecessary radiation. In addition, to minimize the risk of BTI-BSD in patients in radiography, the search for the use or fabrication of a new, lower atomic number material for the bucky table is suggested. Cureus, Inc 2023-02-02 Article PeerReviewed application/pdf en http://irep.iium.edu.my/103580/2/103580_Entrance%20skin%20dose%20%28ESD%29%20and%20bucky%20table.pdf application/pdf en http://irep.iium.edu.my/103580/3/WOS%202023.pdf Sayed, Inayatullah Shah and Roslan, Nurul Shuhada and Syed, Waliullah Shah (2023) Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD). Cureus Journal Medical Science, 15 (2). pp. 1-11. ISSN 2168-8184 https://www.cureus.com/articles/131554-entrance-skin-dose-esd-and-bucky-table-induced-backscattered-dose-bti-bsd-in-abdominal-radiography-with-nanodot-optically-stimulated-luminescence-dosimeter-osld#!/authors 10.7759/cureus.34585
institution Universiti Islam Antarabangsa Malaysia
building IIUM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider International Islamic University Malaysia
content_source IIUM Repository (IREP)
url_provider http://irep.iium.edu.my/
language English
English
topic QC Physics
spellingShingle QC Physics
Sayed, Inayatullah Shah
Roslan, Nurul Shuhada
Syed, Waliullah Shah
Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
description In radiography, inconsistencies in patients' measured entrance skin dose (ESD) exist. There is no published research on the bucky table induced backscattered radiation dose (BTI-BSD). Thus, we aimed to ascertain ESD, calculate the BTI-BSD in abdominal radiography with a nanoDot OSLD, and compare the ESD results with the published data. A Kyoto Kagaku PBU-50 phantom (Kyoto, Japan) in an antero-posterior supine position was exposed, selecting a protocol used for abdominal radiography. The central ray of x-ray beam was pointed at the surface of abdomen at the navel, where a nanoDot dosimeter was placed to measure ESD. For the BTI-BSD, exit dose (ED) was determined by placing a second dosimeter on the exact opposite side (backside) of the phantom from the dosimeter used to determine (ESD) with and without bucky table at identical exposure parameters. The BTI-BSD was calculated as the difference between ED with and without bucky table. The ESD, ED, and BTI-BSD were measured in milligray (mGy). ESD mean values with and without bucky table were 1.97 mGy and 1.84 mGy, whereas ED values were 0.062 mGy and 0.052 mGy, respectively. Results show 2-26% lower ESD values with nanoDot OSLD. The BTI-BSD mean value was found to be approximately 0.01 mGy. A local dose reference level (LDRL) can be established using ESD data to safeguard patients from unnecessary radiation. In addition, to minimize the risk of BTI-BSD in patients in radiography, the search for the use or fabrication of a new, lower atomic number material for the bucky table is suggested.
format Article
author Sayed, Inayatullah Shah
Roslan, Nurul Shuhada
Syed, Waliullah Shah
author_facet Sayed, Inayatullah Shah
Roslan, Nurul Shuhada
Syed, Waliullah Shah
author_sort Sayed, Inayatullah Shah
title Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
title_short Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
title_full Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
title_fullStr Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
title_full_unstemmed Entrance skin dose (ESD) and bucky table induced backscattered dose (BTI-BSD) in abdominal radiography with nanoDot optically stimulated luminescence dosimeter (OSLD)
title_sort entrance skin dose (esd) and bucky table induced backscattered dose (bti-bsd) in abdominal radiography with nanodot optically stimulated luminescence dosimeter (osld)
publisher Cureus, Inc
publishDate 2023
url http://irep.iium.edu.my/103580/2/103580_Entrance%20skin%20dose%20%28ESD%29%20and%20bucky%20table.pdf
http://irep.iium.edu.my/103580/3/WOS%202023.pdf
http://irep.iium.edu.my/103580/
https://www.cureus.com/articles/131554-entrance-skin-dose-esd-and-bucky-table-induced-backscattered-dose-bti-bsd-in-abdominal-radiography-with-nanodot-optically-stimulated-luminescence-dosimeter-osld#!/authors
_version_ 1765296472124817408