Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study

A simple, rapid and sensitive gas chromatographic (GC) method with flame ionization detector (FID) has been developed and validated for simultaneous estimation of camphor, menthol and methyl salicylate (MS). Camphor, menthol and MS were separated at about 2.753, 3.206 and 3.995 min respectively...

Full description

Saved in:
Bibliographic Details
Main Authors: Md Adib, Noor Adibah, Mandal, Uttam Kumar, Mohamed, Farahidah, Chatterjee, Bappaditya
Format: Article
Language:English
English
English
Published: SpringerLink 2017
Subjects:
Online Access:http://irep.iium.edu.my/56687/1/GC%20JPI.pdf
http://irep.iium.edu.my/56687/7/56687_Fast%20and%20simple%20gas%20chromatographic.pdf
http://irep.iium.edu.my/56687/13/56687_Fast%20and%20simple%20gas%20chromatographic%20method_article.pdf
http://irep.iium.edu.my/56687/
https://link.springer.com/article/10.1007%2Fs40005-017-0305-0
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Islam Antarabangsa Malaysia
Language: English
English
English
id my.iium.irep.56687
record_format dspace
spelling my.iium.irep.566872018-03-21T00:47:51Z http://irep.iium.edu.my/56687/ Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study Md Adib, Noor Adibah Mandal, Uttam Kumar Mohamed, Farahidah Chatterjee, Bappaditya RS Pharmacy and materia medica A simple, rapid and sensitive gas chromatographic (GC) method with flame ionization detector (FID) has been developed and validated for simultaneous estimation of camphor, menthol and methyl salicylate (MS). Camphor, menthol and MS were separated at about 2.753, 3.206 and 3.995 min respectively on a capillary column with helium (3.3 ml/min) as carrier gas within 11 min run time. Noninterference of any peak with the peaks of interest confirms the selectivity of method. Derived quantitation limits (QL) were 0.847, 0.684 and 6.507 μg/ml for camphor, menthol and MS respectively. The linear relationship ( R2 > 0.999) between analyte concentration vs detector response was established within a range of QL to 150% of label claim concentration for each analyte. Recovery of each analyte at 50, 100 and 150% of label claim concentration levels were obtained within 99.67–101.53% establishing high accuracy of the method. The method showed acceptable precision with low relative standard deviation or RSD (0.24–1.03%) between percent recoveries for each analyte. RSD for intermediate precision (inter day analysis, analyst variation) was less than 1%. The validated method was successfully applied for quantitative determination of camphor, menthol and MS in stability samples of an analgesic ointment produced by IKOP Sdn. Bhd., Malaysia. SpringerLink 2017-01 Article REM application/pdf en http://irep.iium.edu.my/56687/1/GC%20JPI.pdf application/pdf en http://irep.iium.edu.my/56687/7/56687_Fast%20and%20simple%20gas%20chromatographic.pdf application/pdf en http://irep.iium.edu.my/56687/13/56687_Fast%20and%20simple%20gas%20chromatographic%20method_article.pdf Md Adib, Noor Adibah and Mandal, Uttam Kumar and Mohamed, Farahidah and Chatterjee, Bappaditya (2017) Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study. Journal of Pharmaceutical Investigation. pp. 1-11. ISSN 2093-6214 https://link.springer.com/article/10.1007%2Fs40005-017-0305-0 10.1007/s40005-017-0305-0
institution Universiti Islam Antarabangsa Malaysia
building IIUM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider International Islamic University Malaysia
content_source IIUM Repository (IREP)
url_provider http://irep.iium.edu.my/
language English
English
English
topic RS Pharmacy and materia medica
spellingShingle RS Pharmacy and materia medica
Md Adib, Noor Adibah
Mandal, Uttam Kumar
Mohamed, Farahidah
Chatterjee, Bappaditya
Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
description A simple, rapid and sensitive gas chromatographic (GC) method with flame ionization detector (FID) has been developed and validated for simultaneous estimation of camphor, menthol and methyl salicylate (MS). Camphor, menthol and MS were separated at about 2.753, 3.206 and 3.995 min respectively on a capillary column with helium (3.3 ml/min) as carrier gas within 11 min run time. Noninterference of any peak with the peaks of interest confirms the selectivity of method. Derived quantitation limits (QL) were 0.847, 0.684 and 6.507 μg/ml for camphor, menthol and MS respectively. The linear relationship ( R2 > 0.999) between analyte concentration vs detector response was established within a range of QL to 150% of label claim concentration for each analyte. Recovery of each analyte at 50, 100 and 150% of label claim concentration levels were obtained within 99.67–101.53% establishing high accuracy of the method. The method showed acceptable precision with low relative standard deviation or RSD (0.24–1.03%) between percent recoveries for each analyte. RSD for intermediate precision (inter day analysis, analyst variation) was less than 1%. The validated method was successfully applied for quantitative determination of camphor, menthol and MS in stability samples of an analgesic ointment produced by IKOP Sdn. Bhd., Malaysia.
format Article
author Md Adib, Noor Adibah
Mandal, Uttam Kumar
Mohamed, Farahidah
Chatterjee, Bappaditya
author_facet Md Adib, Noor Adibah
Mandal, Uttam Kumar
Mohamed, Farahidah
Chatterjee, Bappaditya
author_sort Md Adib, Noor Adibah
title Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
title_short Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
title_full Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
title_fullStr Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
title_full_unstemmed Fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
title_sort fast and simple gas chromatographic method for simultaneous estimation of camphor, menthol and methyl salicylate in analgesic ointment: application in stability study
publisher SpringerLink
publishDate 2017
url http://irep.iium.edu.my/56687/1/GC%20JPI.pdf
http://irep.iium.edu.my/56687/7/56687_Fast%20and%20simple%20gas%20chromatographic.pdf
http://irep.iium.edu.my/56687/13/56687_Fast%20and%20simple%20gas%20chromatographic%20method_article.pdf
http://irep.iium.edu.my/56687/
https://link.springer.com/article/10.1007%2Fs40005-017-0305-0
_version_ 1643614975901564928