Modified PET / SPECT cylindrical phantom

It is mandatory to assess / evaluate performance parameters of all medical imaging systems employed in diagnostic imaging modalities before their use in patient studies. For such evaluations a number of commercially available test phantoms exist, e.g., in nuclear medicine tomography - PET/SPECT phan...

Full description

Saved in:
Bibliographic Details
Main Authors: Sayed, Inayatullah Shah, Shah, Ayaz Ali
Other Authors: Magjarevic, R.
Format: Book Chapter
Language:English
English
Published: Springer Berlin Heidelberg 2007
Subjects:
Online Access:http://irep.iium.edu.my/5678/1/Modified_PET-SPECT_Cylindrical_Phantom__IFMBE_WCoMP2006-South_Korea.pdf
http://irep.iium.edu.my/5678/4/bfm_978-3-540-36841-0_1.pdf
http://irep.iium.edu.my/5678/
http://www.springerlink.com/content/j618n362676417xg/
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Islam Antarabangsa Malaysia
Language: English
English
id my.iium.irep.5678
record_format dspace
spelling my.iium.irep.56782013-02-13T07:58:48Z http://irep.iium.edu.my/5678/ Modified PET / SPECT cylindrical phantom Sayed, Inayatullah Shah Shah, Ayaz Ali QC Physics It is mandatory to assess / evaluate performance parameters of all medical imaging systems employed in diagnostic imaging modalities before their use in patient studies. For such evaluations a number of commercially available test phantoms exist, e.g., in nuclear medicine tomography - PET/SPECT phantoms. A cylindrical PET/SPECT phantom (R A Carlson) which consists of hot, cold and linearity inserts has been modified. An additional pair of holes 30 mm diameter (hot regions in a cold background) — including “V” shaped eight pairs of various sized holes — near the edges to the hot regions insert opposite to each other at 180° from the centre of the insert is drilled. For cold regions insert, solid plastic rods 5 cm long (each) with same diameters and exactly the same arrangement as hot region holes pairs is designed. Both inserts have been attached with each other by aligning hot and cold regions for same cross sectional views. The data acquired by this arrangement provides a simple way for measurement and comparison of the performance of a system i.e., image resolution in terms of positions of hot and cold regions in the phantom, lesion detectability. Also the accuracy of scatter and attenuation correction techniques in nuclear medicine tomography may be tested. Springer Berlin Heidelberg Magjarevic, R. Nagel, J. H. 2007 Book Chapter REM application/pdf en http://irep.iium.edu.my/5678/1/Modified_PET-SPECT_Cylindrical_Phantom__IFMBE_WCoMP2006-South_Korea.pdf application/pdf en http://irep.iium.edu.my/5678/4/bfm_978-3-540-36841-0_1.pdf Sayed, Inayatullah Shah and Shah, Ayaz Ali (2007) Modified PET / SPECT cylindrical phantom. In: World Congress on Medical Physics and Biomedical Engineering 2006 : August 27 – September 1, 2006 COEX Seoul, Korea “Imaging the Future Medicine”. IFMBE Proceedings, 14 (11). Springer Berlin Heidelberg, Germany, pp. 1-586. ISBN 978-3-540-36839-7 http://www.springerlink.com/content/j618n362676417xg/ 10.1007/978-3-540-36841-0_414
institution Universiti Islam Antarabangsa Malaysia
building IIUM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider International Islamic University Malaysia
content_source IIUM Repository (IREP)
url_provider http://irep.iium.edu.my/
language English
English
topic QC Physics
spellingShingle QC Physics
Sayed, Inayatullah Shah
Shah, Ayaz Ali
Modified PET / SPECT cylindrical phantom
description It is mandatory to assess / evaluate performance parameters of all medical imaging systems employed in diagnostic imaging modalities before their use in patient studies. For such evaluations a number of commercially available test phantoms exist, e.g., in nuclear medicine tomography - PET/SPECT phantoms. A cylindrical PET/SPECT phantom (R A Carlson) which consists of hot, cold and linearity inserts has been modified. An additional pair of holes 30 mm diameter (hot regions in a cold background) — including “V” shaped eight pairs of various sized holes — near the edges to the hot regions insert opposite to each other at 180° from the centre of the insert is drilled. For cold regions insert, solid plastic rods 5 cm long (each) with same diameters and exactly the same arrangement as hot region holes pairs is designed. Both inserts have been attached with each other by aligning hot and cold regions for same cross sectional views. The data acquired by this arrangement provides a simple way for measurement and comparison of the performance of a system i.e., image resolution in terms of positions of hot and cold regions in the phantom, lesion detectability. Also the accuracy of scatter and attenuation correction techniques in nuclear medicine tomography may be tested.
author2 Magjarevic, R.
author_facet Magjarevic, R.
Sayed, Inayatullah Shah
Shah, Ayaz Ali
format Book Chapter
author Sayed, Inayatullah Shah
Shah, Ayaz Ali
author_sort Sayed, Inayatullah Shah
title Modified PET / SPECT cylindrical phantom
title_short Modified PET / SPECT cylindrical phantom
title_full Modified PET / SPECT cylindrical phantom
title_fullStr Modified PET / SPECT cylindrical phantom
title_full_unstemmed Modified PET / SPECT cylindrical phantom
title_sort modified pet / spect cylindrical phantom
publisher Springer Berlin Heidelberg
publishDate 2007
url http://irep.iium.edu.my/5678/1/Modified_PET-SPECT_Cylindrical_Phantom__IFMBE_WCoMP2006-South_Korea.pdf
http://irep.iium.edu.my/5678/4/bfm_978-3-540-36841-0_1.pdf
http://irep.iium.edu.my/5678/
http://www.springerlink.com/content/j618n362676417xg/
_version_ 1643605594449379328