The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli
The use of maximum length sequence (MLS) stimuli to elicit an auditory brainstem response (ABR) has been limited, in part, by the observation that these stimuli reduce ABR wave amplitudes. This study recorded ABR waveforms from 14 normally hearing adults using MLS click stimuli (maximum stimulu...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
S. Karger AG
2008
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/6545/2/Dzulkarnain2007EffectElectMontOnAmpWaveVABRMLS.pdf http://irep.iium.edu.my/6545/ http://dx.doi.org/10.1159/000107432 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Islam Antarabangsa Malaysia |
Language: | English |
id |
my.iium.irep.6545 |
---|---|
record_format |
dspace |
spelling |
my.iium.irep.65452022-04-12T02:50:21Z http://irep.iium.edu.my/6545/ The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli Dzulkarnain, Ahmad Aidil Arafat Wilson, Wayne J. Bradley, Andrew P. Petoe, Matthew R Medicine (General) The use of maximum length sequence (MLS) stimuli to elicit an auditory brainstem response (ABR) has been limited, in part, by the observation that these stimuli reduce ABR wave amplitudes. This study recorded ABR waveforms from 14 normally hearing adults using MLS click stimuli (maximum stimulus rate = 250 clicks per second) at stimulus levels of 70, 60, 50, 40, 30 and 20 dB nHL, with a vertical and then an ipsilateral electrode montage. The vertical electrode montage produced significantly larger (p ! 0.05) wave V amplitudes, with no change in wave V latencies (p 1 0.05), at all stimulus levels. This result suggests a vertical electrode montage could be used to counter some of the loss in wave V amplitude observed when using MLS stimuli. S. Karger AG 2008 Article PeerReviewed application/pdf en http://irep.iium.edu.my/6545/2/Dzulkarnain2007EffectElectMontOnAmpWaveVABRMLS.pdf Dzulkarnain, Ahmad Aidil Arafat and Wilson, Wayne J. and Bradley, Andrew P. and Petoe, Matthew (2008) The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli. Audiology and Neurotology, 13 (1). pp. 7-12. ISSN 1420-3030 http://dx.doi.org/10.1159/000107432 doi:10.1159/000107432 |
institution |
Universiti Islam Antarabangsa Malaysia |
building |
IIUM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
International Islamic University Malaysia |
content_source |
IIUM Repository (IREP) |
url_provider |
http://irep.iium.edu.my/ |
language |
English |
topic |
R Medicine (General) |
spellingShingle |
R Medicine (General) Dzulkarnain, Ahmad Aidil Arafat Wilson, Wayne J. Bradley, Andrew P. Petoe, Matthew The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
description |
The use of maximum length sequence (MLS) stimuli to elicit
an auditory brainstem response (ABR) has been limited, in
part, by the observation that these stimuli reduce ABR wave
amplitudes. This study recorded ABR waveforms from 14
normally hearing adults using MLS click stimuli (maximum
stimulus rate = 250 clicks per second) at stimulus levels of 70,
60, 50, 40, 30 and 20 dB nHL, with a vertical and then an ipsilateral
electrode montage. The vertical electrode montage
produced significantly larger (p ! 0.05) wave V amplitudes,
with no change in wave V latencies (p 1 0.05), at all stimulus
levels. This result suggests a vertical electrode montage
could be used to counter some of the loss in wave V amplitude
observed when using MLS stimuli. |
format |
Article |
author |
Dzulkarnain, Ahmad Aidil Arafat Wilson, Wayne J. Bradley, Andrew P. Petoe, Matthew |
author_facet |
Dzulkarnain, Ahmad Aidil Arafat Wilson, Wayne J. Bradley, Andrew P. Petoe, Matthew |
author_sort |
Dzulkarnain, Ahmad Aidil Arafat |
title |
The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
title_short |
The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
title_full |
The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
title_fullStr |
The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
title_full_unstemmed |
The Effects of Electrode Montage on the Amplitude of Wave V in the Auditory Brainstem Response to Maximum Length Sequence Stimuli |
title_sort |
effects of electrode montage on the amplitude of wave v in the auditory brainstem response to maximum length sequence stimuli |
publisher |
S. Karger AG |
publishDate |
2008 |
url |
http://irep.iium.edu.my/6545/2/Dzulkarnain2007EffectElectMontOnAmpWaveVABRMLS.pdf http://irep.iium.edu.my/6545/ http://dx.doi.org/10.1159/000107432 |
_version_ |
1731225358908260352 |