Effect of focusing on sizing of side drilled bole (SDH) of different depth using various aperture size of PAUT ultrasonic probe / Wahiddah Long Mohd Nor Azizi
This study used the advance instrument of PAUT, the OMNIScan MX2 provided by Malaysian Nuclear Agency to evaluate the size of SDH by using different focusing and various aperture size of a probe. The type of probe used is 5L64 and the specimen is the lOW Block that consist of Side Drilled Hole (...
Saved in:
Main Author: | |
---|---|
Format: | Student Project |
Language: | English |
Published: |
Faculty Of Applied Sciences
2017
|
Subjects: | |
Online Access: | http://ir.uitm.edu.my/id/eprint/23644/1/PPb_WAHIDDAH%20LONG%20MOHD%20NOR%20AZIZI%20AS%20C%2017_5.PDF http://ir.uitm.edu.my/id/eprint/23644/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Teknologi Mara |
Language: | English |
Summary: | This study used the advance instrument of PAUT, the OMNIScan MX2
provided by Malaysian Nuclear Agency to evaluate the size of SDH by using
different focusing and various aperture size of a probe. The type of probe used
is 5L64 and the specimen is the lOW Block that consist of Side Drilled Hole
(SDH) at different depths. The size of SDH is scanned by varying the parameter
of the instrument at 16 elements and 32 elements and focusing depth of 10mm,
45mm and 75mm. The elements used contribute to the probe aperture size and
gives different range of nearfield which are 20.87mm and 83.89mm. The sizes
of SDH are determined by normalization and the beam profiles are constructed.
Size of scanned SDH are compared focused which are 15mm, 25mm, 32mm,
43mm and 60mm. The percentage by the actual size of the hole on the lOW
block and the percentage error is calculated. The beam profiles showed the
depth of SDH where the beam errors of the focused holes are 23.33%,
116.67%, 20.00%, 53.33% and 120.00%. To minimize the percentage error,
bigger aperture size can be used to increases the probe sensitivity and the
focusing depth must be set within the range of the nearfield. The data obtained
from the experiment give a picture on how the aperture size and the focusing
depth that have been set on the instrument correlated to each other and also the
effect of both parameters on the size of the SDH scanned |
---|