Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom

Purpose: This study aimed to evaluate the effects of various computed tomography (CT) acquisition parameters and metal artifacts on CT number measurement for CT thermometry during CT-guided thermal ablation. Methods: The effects of tube voltage (100–140 kVp), tube current (20–250 mAs), pitch (0.6–1....

Full description

Saved in:
Bibliographic Details
Main Authors: Tan, Daryl, Mohamad, Nurul Ashikin, Wong, Yin How, Yeong, Chai Hong, Cheah, Peng Loon, Sulaiman, Norshazriman, Abdullah, Basri Johan Jeet, Fabell, Mohd Kamil, Lim, Kok Sing
Format: Article
Published: Taylor & Francis 2019
Subjects:
Online Access:http://eprints.um.edu.my/23311/
https://doi.org/10.1080/02656736.2019.1610800
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Malaya
id my.um.eprints.23311
record_format eprints
spelling my.um.eprints.233112020-01-06T02:27:53Z http://eprints.um.edu.my/23311/ Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom Tan, Daryl Mohamad, Nurul Ashikin Wong, Yin How Yeong, Chai Hong Cheah, Peng Loon Sulaiman, Norshazriman Abdullah, Basri Johan Jeet Fabell, Mohd Kamil Lim, Kok Sing R Medicine Purpose: This study aimed to evaluate the effects of various computed tomography (CT) acquisition parameters and metal artifacts on CT number measurement for CT thermometry during CT-guided thermal ablation. Methods: The effects of tube voltage (100–140 kVp), tube current (20–250 mAs), pitch (0.6–1.5) and gantry rotation time (0.5, 1.0 s) as well as metal artifacts from a radiofrequency ablation (RFA) needle on CT number were evaluated using liver tissue equivalent polyacrylamide (PAA) phantom. The correlation between CT number and temperature from 37 to 80 °C was studied on PAA phantom using optimum CT acquisition parameters. Results: No statistical significant difference (p > 0.05) was found on CT numbers under the variation of different acquisition parameters for the same temperature setting. On the other hand, the RFA needle has induced metal artifacts on the CT images of up to 8 mm. The CT numbers decreased linearly when the phantom temperature increased from 37 to 80 °C. A linear regression analysis on the CT numbers and temperature suggested that the CT thermal sensitivity was –0.521 ± 0.061 HU/°C (R2 = 0.998). Conclusion: CT thermometry is feasible for temperature assessment during RFA with the current CT technology, which produced a high CT number reproducibility and stable measurement at different CT acquisition parameters. Despite being affected by metal artifacts, the CT-based thermometry could be further developed as a tissue temperature monitoring tool during CT-guided thermal ablation. © 2019, © 2019 The Author(s). Published with license by Taylor & Francis Group, LLC. Taylor & Francis 2019 Article PeerReviewed Tan, Daryl and Mohamad, Nurul Ashikin and Wong, Yin How and Yeong, Chai Hong and Cheah, Peng Loon and Sulaiman, Norshazriman and Abdullah, Basri Johan Jeet and Fabell, Mohd Kamil and Lim, Kok Sing (2019) Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom. International Journal of Hyperthermia, 36 (1). pp. 553-560. ISSN 0265-6736 https://doi.org/10.1080/02656736.2019.1610800 doi:10.1080/02656736.2019.1610800
institution Universiti Malaya
building UM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Malaya
content_source UM Research Repository
url_provider http://eprints.um.edu.my/
topic R Medicine
spellingShingle R Medicine
Tan, Daryl
Mohamad, Nurul Ashikin
Wong, Yin How
Yeong, Chai Hong
Cheah, Peng Loon
Sulaiman, Norshazriman
Abdullah, Basri Johan Jeet
Fabell, Mohd Kamil
Lim, Kok Sing
Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
description Purpose: This study aimed to evaluate the effects of various computed tomography (CT) acquisition parameters and metal artifacts on CT number measurement for CT thermometry during CT-guided thermal ablation. Methods: The effects of tube voltage (100–140 kVp), tube current (20–250 mAs), pitch (0.6–1.5) and gantry rotation time (0.5, 1.0 s) as well as metal artifacts from a radiofrequency ablation (RFA) needle on CT number were evaluated using liver tissue equivalent polyacrylamide (PAA) phantom. The correlation between CT number and temperature from 37 to 80 °C was studied on PAA phantom using optimum CT acquisition parameters. Results: No statistical significant difference (p > 0.05) was found on CT numbers under the variation of different acquisition parameters for the same temperature setting. On the other hand, the RFA needle has induced metal artifacts on the CT images of up to 8 mm. The CT numbers decreased linearly when the phantom temperature increased from 37 to 80 °C. A linear regression analysis on the CT numbers and temperature suggested that the CT thermal sensitivity was –0.521 ± 0.061 HU/°C (R2 = 0.998). Conclusion: CT thermometry is feasible for temperature assessment during RFA with the current CT technology, which produced a high CT number reproducibility and stable measurement at different CT acquisition parameters. Despite being affected by metal artifacts, the CT-based thermometry could be further developed as a tissue temperature monitoring tool during CT-guided thermal ablation. © 2019, © 2019 The Author(s). Published with license by Taylor & Francis Group, LLC.
format Article
author Tan, Daryl
Mohamad, Nurul Ashikin
Wong, Yin How
Yeong, Chai Hong
Cheah, Peng Loon
Sulaiman, Norshazriman
Abdullah, Basri Johan Jeet
Fabell, Mohd Kamil
Lim, Kok Sing
author_facet Tan, Daryl
Mohamad, Nurul Ashikin
Wong, Yin How
Yeong, Chai Hong
Cheah, Peng Loon
Sulaiman, Norshazriman
Abdullah, Basri Johan Jeet
Fabell, Mohd Kamil
Lim, Kok Sing
author_sort Tan, Daryl
title Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
title_short Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
title_full Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
title_fullStr Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
title_full_unstemmed Experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
title_sort experimental assessment on feasibility of computed tomography-based thermometry for radiofrequency ablation on tissue equivalent polyacrylamide phantom
publisher Taylor & Francis
publishDate 2019
url http://eprints.um.edu.my/23311/
https://doi.org/10.1080/02656736.2019.1610800
_version_ 1654960715631427584