Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy

Dose measurement in high dose rate (HDR) brachytherapy treatment is challenging due to the nature of steep dose gradient characteristic of brachytherapy source. MOSkin is a MOSFET based detector which possesses the main advantages of having small active volume and water equivalent packaging as well...

Full description

Saved in:
Bibliographic Details
Main Authors: Ung, Ngie Min, Jamalludin, Zulaikha, Abdul Malik, Rozita, Ho, Gwo Fuang, Poder, Joel, Cutajar, Dean, Rosenfeld, Anatoly
Format: Conference or Workshop Item
Language:English
Subjects:
Online Access:http://eprints.um.edu.my/35251/1/Profesor%20Madya%20Dr.%20Ung%20Ngie%20Min.pdf
http://eprints.um.edu.my/35251/
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Malaya
Language: English
id my.um.eprints.35251
record_format eprints
spelling my.um.eprints.352512022-10-12T04:39:06Z http://eprints.um.edu.my/35251/ Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy Ung, Ngie Min Jamalludin, Zulaikha Abdul Malik, Rozita Ho, Gwo Fuang Poder, Joel Cutajar, Dean Rosenfeld, Anatoly R Medicine (General) Dose measurement in high dose rate (HDR) brachytherapy treatment is challenging due to the nature of steep dose gradient characteristic of brachytherapy source. MOSkin is a MOSFET based detector which possesses the main advantages of having small active volume and water equivalent packaging as well as providing real-time measurement. It therefore has a high potential to be the detector of choice for implementation of in vivo dosimetry (IVD) programme in clinical brachytherapy. In this work, the aim is to evaluate the feasibility and applicability of MOSkin as an in vivo detector for Cobalt-60 (Co-60) brachytherapy dose measurement during brachytherapy of skin and cervical cancers. The MOSkin was first explored for a dosimetric study at lead-tissue interface to measure the dose to the targeted and normal skin regions during skin brachytherapy. Dose quantification with MOSkin performed in three different source-lead-bolus phantom setup arrangements and during treatment delivery, providing dosimetric information of tumour and normal tissue doses evaluation. There was 20% dose enhancement at the target area from source-lead-bolus (SLB) setup while > 25% dose reduction to normal skin was achieved with source-bolus-lead (SBL) setup arrangement. The suitability of the MOSkin detector for rectal dose measurement during cervical intracavitary brachytherapy (ICBT) was then investigated. Verification of classical cervical pear-shaped isodose distribution from ICBT applicator was performed in a custommade gynaecological phantom which resulted in < 3% dose difference between the TPS planned and MOSkin measured dose. This is followed by clinical in vivo rectal dose measurements in 18 brachytherapy insertions with placement of single MOSkin detector, which resulted in percentage dose difference with planned doses ranging from −16.3% to 14.9%. In conclusion, the MOSkin detector has been used for IVD and the results have verified the suitability and applicability of the MOSkin as detector of choice for IVD in HDR brachytherapy using Co-60. Conference or Workshop Item PeerReviewed text en http://eprints.um.edu.my/35251/1/Profesor%20Madya%20Dr.%20Ung%20Ngie%20Min.pdf Ung, Ngie Min and Jamalludin, Zulaikha and Abdul Malik, Rozita and Ho, Gwo Fuang and Poder, Joel and Cutajar, Dean and Rosenfeld, Anatoly Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy. In: The 15th International Symposium on Radiation Physics, 6-10 December 2021, Kuala Lumpur. (Submitted)
institution Universiti Malaya
building UM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Malaya
content_source UM Research Repository
url_provider http://eprints.um.edu.my/
language English
topic R Medicine (General)
spellingShingle R Medicine (General)
Ung, Ngie Min
Jamalludin, Zulaikha
Abdul Malik, Rozita
Ho, Gwo Fuang
Poder, Joel
Cutajar, Dean
Rosenfeld, Anatoly
Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
description Dose measurement in high dose rate (HDR) brachytherapy treatment is challenging due to the nature of steep dose gradient characteristic of brachytherapy source. MOSkin is a MOSFET based detector which possesses the main advantages of having small active volume and water equivalent packaging as well as providing real-time measurement. It therefore has a high potential to be the detector of choice for implementation of in vivo dosimetry (IVD) programme in clinical brachytherapy. In this work, the aim is to evaluate the feasibility and applicability of MOSkin as an in vivo detector for Cobalt-60 (Co-60) brachytherapy dose measurement during brachytherapy of skin and cervical cancers. The MOSkin was first explored for a dosimetric study at lead-tissue interface to measure the dose to the targeted and normal skin regions during skin brachytherapy. Dose quantification with MOSkin performed in three different source-lead-bolus phantom setup arrangements and during treatment delivery, providing dosimetric information of tumour and normal tissue doses evaluation. There was 20% dose enhancement at the target area from source-lead-bolus (SLB) setup while > 25% dose reduction to normal skin was achieved with source-bolus-lead (SBL) setup arrangement. The suitability of the MOSkin detector for rectal dose measurement during cervical intracavitary brachytherapy (ICBT) was then investigated. Verification of classical cervical pear-shaped isodose distribution from ICBT applicator was performed in a custommade gynaecological phantom which resulted in < 3% dose difference between the TPS planned and MOSkin measured dose. This is followed by clinical in vivo rectal dose measurements in 18 brachytherapy insertions with placement of single MOSkin detector, which resulted in percentage dose difference with planned doses ranging from −16.3% to 14.9%. In conclusion, the MOSkin detector has been used for IVD and the results have verified the suitability and applicability of the MOSkin as detector of choice for IVD in HDR brachytherapy using Co-60.
format Conference or Workshop Item
author Ung, Ngie Min
Jamalludin, Zulaikha
Abdul Malik, Rozita
Ho, Gwo Fuang
Poder, Joel
Cutajar, Dean
Rosenfeld, Anatoly
author_facet Ung, Ngie Min
Jamalludin, Zulaikha
Abdul Malik, Rozita
Ho, Gwo Fuang
Poder, Joel
Cutajar, Dean
Rosenfeld, Anatoly
author_sort Ung, Ngie Min
title Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
title_short Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
title_full Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
title_fullStr Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
title_full_unstemmed Applications of MOSkin as a real-time detector for in vivo dosimetry during high-dose rate Cobalt-60 brachytherapy
title_sort applications of moskin as a real-time detector for in vivo dosimetry during high-dose rate cobalt-60 brachytherapy
url http://eprints.um.edu.my/35251/1/Profesor%20Madya%20Dr.%20Ung%20Ngie%20Min.pdf
http://eprints.um.edu.my/35251/
_version_ 1748181066462527488