Deviations in red cell blood component temperature during laboratory processing and at blood return monitored by a time-temperature indicator device

Objectives To evaluate the performance and utility of a time-temperature indicator (TTI) to determine the cumulative exposure time (CET) of red cell components (RCC) to temperatures above 10 degrees C occurring within and outside the transfusion laboratory. Background and Objectives Blood centres of...

Full description

Saved in:
Bibliographic Details
Main Authors: Cheng, Jian Yuan, Samudram, Hemalatha, Lee, Christina Lai Ling, Nadarajan, Veera Sekaran
Format: Article
Published: Wiley 2022
Subjects:
Online Access:http://eprints.um.edu.my/40387/
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Malaya
Description
Summary:Objectives To evaluate the performance and utility of a time-temperature indicator (TTI) to determine the cumulative exposure time (CET) of red cell components (RCC) to temperatures above 10 degrees C occurring within and outside the transfusion laboratory. Background and Objectives Blood centres often use the `30 or 60-min rule' for accepting RCC exposed to room temperature (RT) back into inventory. Effective monitoring of these temperature deviations is however lacking. Materials and Methods A Timestrip PLUS (R) TP153 10 degrees C (TS + 10) TTI was attached to RCC units after preparation of the unit in the blood bank or on issue to the ward, to track the CET > 10 degrees C during laboratory processing and outside the transfusion laboratory. Results The mean CET of 153 RCC tracked within the laboratory was 56 min. Sixty-four (41.8%) and 34 (22.2%) of RCC had core temperature (CT) >10 degrees C for more than 30 and 60 min, respectively. Among the 69 RCC that were returned unused, 27 (39.1%), 17 (24.6%) and 5 (7.2%) RCC units had CT >10 degrees C for more than 30, 60 and 120 min respectively. Conclusion A large proportion of RCC have CT >10 degrees C exceeding 30 min during handling within the transfusion laboratory, as well as when RCC are returned unused from transfusion locations. Corrective measures should be implemented to better manage the cold chain to avoid undesirable consequences to blood transfusion. A temperature sensitive device that can also indicate CET can be employed to objectively monitor the period that RCC remained at a CT that exceeds 10 degrees C.