Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report
The releases of harmful ions from cobalt based alloy to host tissues have raised significant health concerns. Carbon contents in this alloy may influence ions release but has yet investigated. It is hypothesized that carbon contents in this alloy will help the formation of oxide layer during thermal...
Saved in:
Main Authors: | , , , , , , , , |
---|---|
Format: | Article |
Published: |
Elsevier
2014
|
Subjects: | |
Online Access: | http://umpir.ump.edu.my/id/eprint/5380/ http://www.scientific.net/AMR.845.462 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Malaysia Pahang |
id |
my.ump.umpir.5380 |
---|---|
record_format |
eprints |
spelling |
my.ump.umpir.53802018-01-31T00:06:23Z http://umpir.ump.edu.my/id/eprint/5380/ Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report Rosdi, Daud Mas Ayu, Hassan S., Izman Mohammed Rafiq, Abdul Kadir A., Shah Mohd Faiz, Mohd Yusoff M. W., Shamsiah T. M., Yong T., Kamarul TJ Mechanical engineering and machinery The releases of harmful ions from cobalt based alloy to host tissues have raised significant health concerns. Carbon contents in this alloy may influence ions release but has yet investigated. It is hypothesized that carbon contents in this alloy will help the formation of oxide layer during thermal oxidation process and hence reducing the release of Co/Cr ions after implantation. In this study, Co-Cr-Mo alloy with carbon concentrations of 0.03% and 0.24% were oxidized at 1050°C for 3 hours under atmospheric condition. The oxidized substrates were characterized under FESEM and subjected to circulating immersion test in simulated body fluid (SBF) for 21 days. Metal ions release was measured using inductively coupled plasma-mass spectrometry (ICP-MS) at day 0, 7, 14 and 21. Oxidized high carbon samples show denser and a more uniform oxide layer than samples with low carbon contents. It is found that compact oxide structure promotes less metal ions release during immersion. Elsevier 2014 Article PeerReviewed Rosdi, Daud and Mas Ayu, Hassan and S., Izman and Mohammed Rafiq, Abdul Kadir and A., Shah and Mohd Faiz, Mohd Yusoff and M. W., Shamsiah and T. M., Yong and T., Kamarul (2014) Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report. Advanced Materials Research, 845. pp. 462-466. ISSN 1662-8985 http://www.scientific.net/AMR.845.462 DOI:10.4028/www.scientific.net/AMR.845.462 |
institution |
Universiti Malaysia Pahang |
building |
UMP Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Malaysia Pahang |
content_source |
UMP Institutional Repository |
url_provider |
http://umpir.ump.edu.my/ |
topic |
TJ Mechanical engineering and machinery |
spellingShingle |
TJ Mechanical engineering and machinery Rosdi, Daud Mas Ayu, Hassan S., Izman Mohammed Rafiq, Abdul Kadir A., Shah Mohd Faiz, Mohd Yusoff M. W., Shamsiah T. M., Yong T., Kamarul Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
description |
The releases of harmful ions from cobalt based alloy to host tissues have raised significant health concerns. Carbon contents in this alloy may influence ions release but has yet investigated. It is hypothesized that carbon contents in this alloy will help the formation of oxide layer during thermal oxidation process and hence reducing the release of Co/Cr ions after implantation. In this study, Co-Cr-Mo alloy with carbon concentrations of 0.03% and 0.24% were oxidized at 1050°C for 3 hours under atmospheric condition. The oxidized substrates were characterized under FESEM and subjected to circulating immersion test in simulated body fluid (SBF) for 21 days. Metal ions release was measured using inductively coupled plasma-mass spectrometry (ICP-MS) at day 0, 7, 14 and 21. Oxidized high carbon samples show denser and a more uniform oxide layer than samples with low carbon contents. It is found that compact oxide structure promotes less metal ions release during immersion. |
format |
Article |
author |
Rosdi, Daud Mas Ayu, Hassan S., Izman Mohammed Rafiq, Abdul Kadir A., Shah Mohd Faiz, Mohd Yusoff M. W., Shamsiah T. M., Yong T., Kamarul |
author_facet |
Rosdi, Daud Mas Ayu, Hassan S., Izman Mohammed Rafiq, Abdul Kadir A., Shah Mohd Faiz, Mohd Yusoff M. W., Shamsiah T. M., Yong T., Kamarul |
author_sort |
Rosdi, Daud |
title |
Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
title_short |
Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
title_full |
Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
title_fullStr |
Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
title_full_unstemmed |
Influence of Carbon Concentrations in Reducing Co and Cr Ions Release in Cobalt Based Implant: A Preliminary Report |
title_sort |
influence of carbon concentrations in reducing co and cr ions release in cobalt based implant: a preliminary report |
publisher |
Elsevier |
publishDate |
2014 |
url |
http://umpir.ump.edu.my/id/eprint/5380/ http://www.scientific.net/AMR.845.462 |
_version_ |
1643665188556111872 |