Investigate the capability of INAA absolute method to determine the concentrations of 238U and 232Th in rock samples
This work aimed to study the capability of INAA absolute method in determining the elemental concentration of 238U and 232Th in the rock samples. The INAA absolute method was implemented in PUSPATI TRIGA Mark II research reactor, Malaysian Nuclear Agency (NM). The accuracy of INAA absolute meth...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | E-Article |
Language: | English |
Published: |
Springer Netherlands
2014
|
Subjects: | |
Online Access: | http://ir.unimas.my/id/eprint/10266/1/Investigate%20the%20capability%20of%20INAA%20absolute%20method%20to%20determine%20the%20concentrations%20of%20238U%20and%20232Th%20in%20rock%20samples%20%28abstract%29.pdf http://ir.unimas.my/id/eprint/10266/ https://scholar.google.com.my/citations?view_op=view_citation&hl=en&user=KmdSVJMAAAAJ&cstart=20&pagesize=80&citation_for_view=KmdSVJMAAAAJ:MXK_kJrjxJIC |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Malaysia Sarawak |
Language: | English |
Summary: | This work aimed to study the capability of
INAA absolute method in determining the elemental concentration
of 238U and 232Th in the rock samples. The INAA
absolute method was implemented in PUSPATI TRIGA
Mark II research reactor, Malaysian Nuclear Agency (NM).
The accuracy of INAA absolute method was performed by
analyzing the IAEA certified reference material (CRM)
Soil-7. The analytical results showed the deviations between
experimental and certified values were mostly less than
10 % with Z-score in most cases less than 1. In general, the
results of analysed CRM Soil-7 show a good agreement
between certified and experimental results which mean that
the INAA absolute method can be used accurately for elemental
analysis of uranium and thorium in various types of
samples. The concentration of 238U and 232Th ranged from
1.77 to 24.25 and 0.88 to 95.50 ppm respectively. The
highest value of 238U and 232Th was recorded for granite
rock sample G17 of 238U and sample G9 of 232Th, whereas
the lower value was 1.77 ppm of 238U recorded in sandstone
rock and 0.88 ppm of 232Th for gabbro. Moreover, a comparison
of the 238U and 232Th results obtained by the INAA
absolute method shows an acceptable level of consistency
with those obtained by the INAA relative method. |
---|