Estimating local diagnostic reference levels for mammography in Dubai

As the total volume of mammograms in Dubai is increasing consistently, it is crucial to focus on the process of dose optimization by determining dose reference levels for such sensitive radiographic examinations as mammography. This work aimed to determine local diagnostic reference levels (DRLs) fo...

Full description

Saved in:
Bibliographic Details
Main Authors: Abdulwahid Noor, Kaltham, Mohd Norsuddin, Norhashimah, Abdul Karim, Muhammad Khalis, Che Isa, Iza Nurzawani, Alshamsi, Wadha
Format: Article
Published: MDPI AG 2024
Online Access:http://psasir.upm.edu.my/id/eprint/105864/
https://www.mdpi.com/2075-4418/14/1/8
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Putra Malaysia
Description
Summary:As the total volume of mammograms in Dubai is increasing consistently, it is crucial to focus on the process of dose optimization by determining dose reference levels for such sensitive radiographic examinations as mammography. This work aimed to determine local diagnostic reference levels (DRLs) for mammography procedures in Dubai at different ranges of breast thickness. A total of 2599 anonymized mammograms were randomly retrieved from a central dose survey database. Mammographic cases for screening women aged from 40 to 69 years were included, while cases of breast implants and breast thickness outside the range of 20–100 mm were excluded. Mean, median, and 75 percentiles were obtained for the mean glandular dose (MGD) distribution of each mammography projection for all compressed breast thickness (CBT) ranges. The local DRLs for mammography in Dubai were found to be between 0.80 mGy and 0.82 mGy for the craniocaudal (CC) projection and between 0.89 mGy and 0.971.8 mGy for the mediolateral oblique (MLO) projection. Local DRLs were proposed according to different breast thicknesses, starting from 20 to 100 mm. All groups of CBT showed a slight difference in MGD values, with higher values in MLO views rather than CC views. The local DRLs in this study were lower than some other Middle Eastern countries and lower than the standard reference levels reported by the International Atomic Energy Agency (IAEA) at 3 mGy/view.