An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer

The performance of an Inductively Coupled Plasma (ICP) emission spectrometer was evaluated in' terms of its monochromator resettability, ICP generator stability and detection limits. The studies showed that the analytical system exhibited similar performance to simultaneous direct reading spe...

Full description

Saved in:
Bibliographic Details
Main Authors: Lee, Chnoong Kheng, Hoh, R., Low, Kun She
Format: Article
Language:English
English
Published: 1982
Online Access:http://psasir.upm.edu.my/id/eprint/2105/1/An_Evaluation_of_An_Inductively_Coupled_Plasma_%28ICP%29.pdf
http://psasir.upm.edu.my/id/eprint/2105/
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Putra Malaysia
Language: English
English
id my.upm.eprints.2105
record_format eprints
spelling my.upm.eprints.21052013-05-27T06:58:51Z http://psasir.upm.edu.my/id/eprint/2105/ An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer Lee, Chnoong Kheng Hoh, R. Low, Kun She The performance of an Inductively Coupled Plasma (ICP) emission spectrometer was evaluated in' terms of its monochromator resettability, ICP generator stability and detection limits. The studies showed that the analytical system exhibited similar performance to simultaneous direct reading spectrometers. The reliability of the monochromator, torch heigh t adjustor and the stability of the source unit were confirmed. 1982 Article NonPeerReviewed application/pdf en http://psasir.upm.edu.my/id/eprint/2105/1/An_Evaluation_of_An_Inductively_Coupled_Plasma_%28ICP%29.pdf Lee, Chnoong Kheng and Hoh, R. and Low, Kun She (1982) An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer. Pertanika, 5 (1). pp. 39-44. English
institution Universiti Putra Malaysia
building UPM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Putra Malaysia
content_source UPM Institutional Repository
url_provider http://psasir.upm.edu.my/
language English
English
description The performance of an Inductively Coupled Plasma (ICP) emission spectrometer was evaluated in' terms of its monochromator resettability, ICP generator stability and detection limits. The studies showed that the analytical system exhibited similar performance to simultaneous direct reading spectrometers. The reliability of the monochromator, torch heigh t adjustor and the stability of the source unit were confirmed.
format Article
author Lee, Chnoong Kheng
Hoh, R.
Low, Kun She
spellingShingle Lee, Chnoong Kheng
Hoh, R.
Low, Kun She
An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
author_facet Lee, Chnoong Kheng
Hoh, R.
Low, Kun She
author_sort Lee, Chnoong Kheng
title An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
title_short An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
title_full An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
title_fullStr An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
title_full_unstemmed An Evaluation of an Inductively Coupled Plasma (ICP) Emission Spectrometer
title_sort evaluation of an inductively coupled plasma (icp) emission spectrometer
publishDate 1982
url http://psasir.upm.edu.my/id/eprint/2105/1/An_Evaluation_of_An_Inductively_Coupled_Plasma_%28ICP%29.pdf
http://psasir.upm.edu.my/id/eprint/2105/
_version_ 1643822219636244480