Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method
This study presents a simple and non-destructive procedure to determine ethanol concentration of ethanol/water solution at ambient temperature based on HP85070B open ended coaxial dielectric constant measurement. The motivation for this study stems from the fact that the `true' concentration of...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Conference or Workshop Item |
Language: | English |
Published: |
IEEE
2016
|
Online Access: | http://psasir.upm.edu.my/id/eprint/59436/1/Determination%20of%20ethanol%20concentration%20of%20ethanolwater%20mixture%20solutions%20with%20open%20ended%20coaxial%20method.pdf http://psasir.upm.edu.my/id/eprint/59436/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Putra Malaysia |
Language: | English |
id |
my.upm.eprints.59436 |
---|---|
record_format |
eprints |
spelling |
my.upm.eprints.594362018-03-05T02:32:56Z http://psasir.upm.edu.my/id/eprint/59436/ Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method Khalid, Norashikin Zainuddin, Mohamad Faiz Abbas, Zulkifly Mohamed, Tity Nazleen Sabli, Nordin This study presents a simple and non-destructive procedure to determine ethanol concentration of ethanol/water solution at ambient temperature based on HP85070B open ended coaxial dielectric constant measurement. The motivation for this study stems from the fact that the `true' concentration of ethanol in a `labeled' container could be `spoiled' due to hygroscopic nature of ethanol, evaporation and other factors. The dielectric constant measurements of eight samples with different molar fraction were validated with modified Cole-Cole-Debye (3CC) simulations. A 5th degree polynomial calibration equation was developed based on 3CC simulations with R2 = 0.9998 and used to estimate the `true' concentration of three ethanol samples obtained from `labeled' containers kept in laboratory. The ethanol concentrations of two of the samples were found to be much lower than what was indicated on their labels, hence, assumed spoiled. IEEE 2016 Conference or Workshop Item PeerReviewed text en http://psasir.upm.edu.my/id/eprint/59436/1/Determination%20of%20ethanol%20concentration%20of%20ethanolwater%20mixture%20solutions%20with%20open%20ended%20coaxial%20method.pdf Khalid, Norashikin and Zainuddin, Mohamad Faiz and Abbas, Zulkifly and Mohamed, Tity Nazleen and Sabli, Nordin (2016) Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method. In: 2016 International Conference on Advances in Electrical, Electronic and Systems Engineering (ICAEES), 14-16 Nov. 2016, Putrajaya, Malaysia. (pp. 652-656). 10.1109/ICAEES.2016.7888127 |
institution |
Universiti Putra Malaysia |
building |
UPM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Putra Malaysia |
content_source |
UPM Institutional Repository |
url_provider |
http://psasir.upm.edu.my/ |
language |
English |
description |
This study presents a simple and non-destructive procedure to determine ethanol concentration of ethanol/water solution at ambient temperature based on HP85070B open ended coaxial dielectric constant measurement. The motivation for this study stems from the fact that the `true' concentration of ethanol in a `labeled' container could be `spoiled' due to hygroscopic nature of ethanol, evaporation and other factors. The dielectric constant measurements of eight samples with different molar fraction were validated with modified Cole-Cole-Debye (3CC) simulations. A 5th degree polynomial calibration equation was developed based on 3CC simulations with R2 = 0.9998 and used to estimate the `true' concentration of three ethanol samples obtained from `labeled' containers kept in laboratory. The ethanol concentrations of two of the samples were found to be much lower than what was indicated on their labels, hence, assumed spoiled. |
format |
Conference or Workshop Item |
author |
Khalid, Norashikin Zainuddin, Mohamad Faiz Abbas, Zulkifly Mohamed, Tity Nazleen Sabli, Nordin |
spellingShingle |
Khalid, Norashikin Zainuddin, Mohamad Faiz Abbas, Zulkifly Mohamed, Tity Nazleen Sabli, Nordin Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
author_facet |
Khalid, Norashikin Zainuddin, Mohamad Faiz Abbas, Zulkifly Mohamed, Tity Nazleen Sabli, Nordin |
author_sort |
Khalid, Norashikin |
title |
Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
title_short |
Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
title_full |
Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
title_fullStr |
Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
title_full_unstemmed |
Determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
title_sort |
determination of ethanol concentration of ethanol/water mixture solutions with open ended coaxial method |
publisher |
IEEE |
publishDate |
2016 |
url |
http://psasir.upm.edu.my/id/eprint/59436/1/Determination%20of%20ethanol%20concentration%20of%20ethanolwater%20mixture%20solutions%20with%20open%20ended%20coaxial%20method.pdf http://psasir.upm.edu.my/id/eprint/59436/ |
_version_ |
1643837071352135680 |