Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle
A portable ion-selective electrode (ISE) meter (LAQUAtwin B-722; Horiba Instruments Inc., Irvine, CA) is available for measuring the sodium ion concentration ([Na]) in biological fluids. The objective of this study was to characterize the analytical performance of the ISE meter in measuring [Na] in...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Published: |
Elsevier
2019
|
Online Access: | http://psasir.upm.edu.my/id/eprint/79772/ https://www.sciencedirect.com/science/article/pii/S0022030219305272 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Putra Malaysia |
id |
my.upm.eprints.79772 |
---|---|
record_format |
eprints |
spelling |
my.upm.eprints.797722022-11-04T09:36:48Z http://psasir.upm.edu.my/id/eprint/79772/ Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle A. A., Megahed Hiew, M. Grünberg, W. Trefz, F. M. Constable, P. D. A portable ion-selective electrode (ISE) meter (LAQUAtwin B-722; Horiba Instruments Inc., Irvine, CA) is available for measuring the sodium ion concentration ([Na]) in biological fluids. The objective of this study was to characterize the analytical performance of the ISE meter in measuring [Na] in whole-blood, plasma, milk, abomasal fluid, and urine samples from cattle. Method comparison studies were performed using whole-blood and plasma samples from 106 sick calves and 11 sick cows admitted to a veterinary teaching hospital, 80 milk and 206 urine samples from 16 lactating Holstein-Friesian cows with experimentally induced free water, electrolyte, and acid-base imbalances, and 67 abomasal fluid samples from 7 healthy male Holstein-Friesian calves fed fresh milk with or without an oral electrolyte solution. Deming regression and Bland-Altman plots were used to determine the accuracy of the meter against reference methods. The meter used in direct mode on undiluted samples measured whole-blood [Na] 9.7 mmol/L (7.3%) lower than a direct ISE reference method and plasma [Na] 16.7 mmol/L (12.7%) lower than an indirect ISE reference method. The meter run in direct mode measured milk [Na] 3.1 mmol/L lower and abomasal fluid [Na] 9.0% lower than indirect ISE reference methods. The meter run in indirect mode on diluted samples accurately measured urine [Na] compared with an indirect ISE reference method. We conclude that, after adjustment for the bias determined from Bland-Altman plots, the LAQUAtwin ISE meter provides a clinically useful and low-cost cow-side instrument for measuring [Na] in whole blood, plasma, milk, and abomasal fluid. Elsevier 2019 Article PeerReviewed A. A., Megahed and Hiew, M. and Grünberg, W. and Trefz, F. M. and Constable, P. D. (2019) Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle. Journal of Dairy Science, 102 (8). pp. 7435-7444. ISSN 0022-0302; ESSN: 1525-3198 https://www.sciencedirect.com/science/article/pii/S0022030219305272 10.3168/jds.2018-16198 |
institution |
Universiti Putra Malaysia |
building |
UPM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Putra Malaysia |
content_source |
UPM Institutional Repository |
url_provider |
http://psasir.upm.edu.my/ |
description |
A portable ion-selective electrode (ISE) meter (LAQUAtwin B-722; Horiba Instruments Inc., Irvine, CA) is available for measuring the sodium ion concentration ([Na]) in biological fluids. The objective of this study was to characterize the analytical performance of the ISE meter in measuring [Na] in whole-blood, plasma, milk, abomasal fluid, and urine samples from cattle. Method comparison studies were performed using whole-blood and plasma samples from 106 sick calves and 11 sick cows admitted to a veterinary teaching hospital, 80 milk and 206 urine samples from 16 lactating Holstein-Friesian cows with experimentally induced free water, electrolyte, and acid-base imbalances, and 67 abomasal fluid samples from 7 healthy male Holstein-Friesian calves fed fresh milk with or without an oral electrolyte solution. Deming regression and Bland-Altman plots were used to determine the accuracy of the meter against reference methods. The meter used in direct mode on undiluted samples measured whole-blood [Na] 9.7 mmol/L (7.3%) lower than a direct ISE reference method and plasma [Na] 16.7 mmol/L (12.7%) lower than an indirect ISE reference method. The meter run in direct mode measured milk [Na] 3.1 mmol/L lower and abomasal fluid [Na] 9.0% lower than indirect ISE reference methods. The meter run in indirect mode on diluted samples accurately measured urine [Na] compared with an indirect ISE reference method. We conclude that, after adjustment for the bias determined from Bland-Altman plots, the LAQUAtwin ISE meter provides a clinically useful and low-cost cow-side instrument for measuring [Na] in whole blood, plasma, milk, and abomasal fluid. |
format |
Article |
author |
A. A., Megahed Hiew, M. Grünberg, W. Trefz, F. M. Constable, P. D. |
spellingShingle |
A. A., Megahed Hiew, M. Grünberg, W. Trefz, F. M. Constable, P. D. Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
author_facet |
A. A., Megahed Hiew, M. Grünberg, W. Trefz, F. M. Constable, P. D. |
author_sort |
A. A., Megahed |
title |
Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
title_short |
Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
title_full |
Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
title_fullStr |
Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
title_full_unstemmed |
Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
title_sort |
evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle |
publisher |
Elsevier |
publishDate |
2019 |
url |
http://psasir.upm.edu.my/id/eprint/79772/ https://www.sciencedirect.com/science/article/pii/S0022030219305272 |
_version_ |
1751538198283026432 |