Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera

The aim of this study is to evaluate the effects of energy window width on the planar image quality of thyroid phantom. The values of contrast and signal to noise ratio (SNR) were evaluated in this study. It was carried out by using a thyroid phantom that had been filled with 5.566 mCi Tc-99m rad...

Full description

Saved in:
Bibliographic Details
Main Author: Jamsari, Nur Liyana
Format: Monograph
Language:English
Published: Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2015
Subjects:
Online Access:http://eprints.usm.my/58074/1/NUR%20LIYANA%20BINTI%20JAMSARI-24%20pages.pdf
http://eprints.usm.my/58074/
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Sains Malaysia
Language: English
id my.usm.eprints.58074
record_format eprints
spelling my.usm.eprints.58074 http://eprints.usm.my/58074/ Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera Jamsari, Nur Liyana TK1-9971 Electrical engineering. Electronics. Nuclear engineering The aim of this study is to evaluate the effects of energy window width on the planar image quality of thyroid phantom. The values of contrast and signal to noise ratio (SNR) were evaluated in this study. It was carried out by using a thyroid phantom that had been filled with 5.566 mCi Tc-99m radioactive sources. A static image of the thyroid phantom was captured using the single headed gamma camera (Symbia-E) at five different energy window widths of 10%, 15%, 20%, 30% and 40% which was equipped with LEHR collimator, stopping condition 300 kilocounts and matrix size of 128 x 128. The average counts obtained were recorded in a table and the contrast and SNR for hot and cold nodules were calculated using the given equation. For contrast, 15% energy window width gave the highest values for the hot nodules but for the cold nodules, it gave lower contrast values compared to 40%. However, the cold nodule at 15% energy window width gave the higher contrast values than 10%, 20% and 30% energy window widths. For SNR, the energy window width of 15% gave the highest values at all four nodules (two hot nodules and two cold nodules). Therefore, the optimum energy window width was the 15% for Tc-99m radioactive and planar imaging for thyroid. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2015 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/58074/1/NUR%20LIYANA%20BINTI%20JAMSARI-24%20pages.pdf Jamsari, Nur Liyana (2015) Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera. Project Report. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia. (Submitted)
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic TK1-9971 Electrical engineering. Electronics. Nuclear engineering
spellingShingle TK1-9971 Electrical engineering. Electronics. Nuclear engineering
Jamsari, Nur Liyana
Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
description The aim of this study is to evaluate the effects of energy window width on the planar image quality of thyroid phantom. The values of contrast and signal to noise ratio (SNR) were evaluated in this study. It was carried out by using a thyroid phantom that had been filled with 5.566 mCi Tc-99m radioactive sources. A static image of the thyroid phantom was captured using the single headed gamma camera (Symbia-E) at five different energy window widths of 10%, 15%, 20%, 30% and 40% which was equipped with LEHR collimator, stopping condition 300 kilocounts and matrix size of 128 x 128. The average counts obtained were recorded in a table and the contrast and SNR for hot and cold nodules were calculated using the given equation. For contrast, 15% energy window width gave the highest values for the hot nodules but for the cold nodules, it gave lower contrast values compared to 40%. However, the cold nodule at 15% energy window width gave the higher contrast values than 10%, 20% and 30% energy window widths. For SNR, the energy window width of 15% gave the highest values at all four nodules (two hot nodules and two cold nodules). Therefore, the optimum energy window width was the 15% for Tc-99m radioactive and planar imaging for thyroid.
format Monograph
author Jamsari, Nur Liyana
author_facet Jamsari, Nur Liyana
author_sort Jamsari, Nur Liyana
title Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
title_short Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
title_full Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
title_fullStr Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
title_full_unstemmed Effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
title_sort effects of energy window width on planar image quality of thyroid imaging using single-headed gamma camera
publisher Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia
publishDate 2015
url http://eprints.usm.my/58074/1/NUR%20LIYANA%20BINTI%20JAMSARI-24%20pages.pdf
http://eprints.usm.my/58074/
_version_ 1765297642909204480