Comparative study of micromixers for laminar blood mixing
Miniaturization is the trend in analytical chemistry and life science. It has been emerging into the research field of microfluidics in the application of LOC. The application is used for biochemistry analysis and require a rapid mixing in small area. Due to laminar flow (Reynold Number < 1) p...
Saved in:
Main Author: | |
---|---|
Format: | Thesis |
Language: | English English English |
Published: |
2013
|
Subjects: | |
Online Access: | http://eprints.uthm.edu.my/6658/1/24p%20KHAIRUL%20NADIAH%20KHALID.pdf http://eprints.uthm.edu.my/6658/2/KHAIRUL%20NADIAH%20KHALID%20COPYRIGHT%20DECLARATION.pdf http://eprints.uthm.edu.my/6658/3/KHAIRUL%20NADIAH%20KHALID%20WATERMARK.pdf http://eprints.uthm.edu.my/6658/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Tun Hussein Onn Malaysia |
Language: | English English English |
id |
my.uthm.eprints.6658 |
---|---|
record_format |
eprints |
spelling |
my.uthm.eprints.66582022-03-14T01:36:21Z http://eprints.uthm.edu.my/6658/ Comparative study of micromixers for laminar blood mixing Khalid, Khairul Nadiah TA164 Bioengineering Miniaturization is the trend in analytical chemistry and life science. It has been emerging into the research field of microfluidics in the application of LOC. The application is used for biochemistry analysis and require a rapid mixing in small area. Due to laminar flow (Reynold Number < 1) passive micromixer is the best method in fluids mixing. Passive micromixer also depend on the channel geometry for mixing effectiveness. In this study, four different micromixers were evaluated based on the baseline control Y-micromixer. The micromixers are internal rib micromixer, patterned grooves micromixer, obstruction micromixer and slanted rib micromixer. These micromixer has 1000μm channel length, 150μm inlet length, 90o between inlets ports, width and depth are 40μm each. The fluids used for mixing were blood which has 1.0 × 10-6 kg/μms of viscosity and toluene which has low viscosity than blood (0.664 × 10-9 kg/μms). The fluids used to evaluate the differences in term of their visual performance based image’s standard deviation by plotting the graph and mixing efficiency by calculation. Based on these evaluations, the slanted rib micromixer is the best micromixer design with the highest mixing efficiency of 99.85% at the outlet of the channel. 2013-07 Thesis NonPeerReviewed text en http://eprints.uthm.edu.my/6658/1/24p%20KHAIRUL%20NADIAH%20KHALID.pdf text en http://eprints.uthm.edu.my/6658/2/KHAIRUL%20NADIAH%20KHALID%20COPYRIGHT%20DECLARATION.pdf text en http://eprints.uthm.edu.my/6658/3/KHAIRUL%20NADIAH%20KHALID%20WATERMARK.pdf Khalid, Khairul Nadiah (2013) Comparative study of micromixers for laminar blood mixing. Masters thesis, Universiti Tun Hussein Malaysia. |
institution |
Universiti Tun Hussein Onn Malaysia |
building |
UTHM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Tun Hussein Onn Malaysia |
content_source |
UTHM Institutional Repository |
url_provider |
http://eprints.uthm.edu.my/ |
language |
English English English |
topic |
TA164 Bioengineering |
spellingShingle |
TA164 Bioengineering Khalid, Khairul Nadiah Comparative study of micromixers for laminar blood mixing |
description |
Miniaturization is the trend in analytical chemistry and life science. It has been
emerging into the research field of microfluidics in the application of LOC. The
application is used for biochemistry analysis and require a rapid mixing in small area.
Due to laminar flow (Reynold Number < 1) passive micromixer is the best method in
fluids mixing. Passive micromixer also depend on the channel geometry for mixing
effectiveness. In this study, four different micromixers were evaluated based on the
baseline control Y-micromixer. The micromixers are internal rib micromixer, patterned
grooves micromixer, obstruction micromixer and slanted rib micromixer. These
micromixer has 1000μm channel length, 150μm inlet length, 90o between inlets ports,
width and depth are 40μm each. The fluids used for mixing were blood which has 1.0 ×
10-6 kg/μms of viscosity and toluene which has low viscosity than blood (0.664 × 10-9
kg/μms). The fluids used to evaluate the differences in term of their visual performance
based image’s standard deviation by plotting the graph and mixing efficiency by
calculation. Based on these evaluations, the slanted rib micromixer is the best
micromixer design with the highest mixing efficiency of 99.85% at the outlet of the
channel. |
format |
Thesis |
author |
Khalid, Khairul Nadiah |
author_facet |
Khalid, Khairul Nadiah |
author_sort |
Khalid, Khairul Nadiah |
title |
Comparative study of micromixers for laminar blood mixing |
title_short |
Comparative study of micromixers for laminar blood mixing |
title_full |
Comparative study of micromixers for laminar blood mixing |
title_fullStr |
Comparative study of micromixers for laminar blood mixing |
title_full_unstemmed |
Comparative study of micromixers for laminar blood mixing |
title_sort |
comparative study of micromixers for laminar blood mixing |
publishDate |
2013 |
url |
http://eprints.uthm.edu.my/6658/1/24p%20KHAIRUL%20NADIAH%20KHALID.pdf http://eprints.uthm.edu.my/6658/2/KHAIRUL%20NADIAH%20KHALID%20COPYRIGHT%20DECLARATION.pdf http://eprints.uthm.edu.my/6658/3/KHAIRUL%20NADIAH%20KHALID%20WATERMARK.pdf http://eprints.uthm.edu.my/6658/ |
_version_ |
1738581518686617600 |