In silico of the different pin size of unilateral tibia external fixator: a pilot study

External fixator is the promising medical device that can be used for treating bone fracture of tibia. The healing process can be taken place if some parameters are considered such as biomechanical aspect. One of them is the diameter of pin. However, the information regarding the biomechanics charac...

Full description

Saved in:
Bibliographic Details
Main Authors: Zainudin, Nur Amalina, Ramlee, Muhammad Hanif
Format: Article
Language:English
Published: ECronicon 2019
Subjects:
Online Access:http://eprints.utm.my/id/eprint/88086/1/MuhammadHanifRamlee2019_InSilicooftheDifferentPinSize.pdf
http://eprints.utm.my/id/eprint/88086/
https://www.ecronicon.com/ecor/pdf/ECOR-10-00512.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Teknologi Malaysia
Language: English
id my.utm.88086
record_format eprints
spelling my.utm.880862020-12-15T00:12:35Z http://eprints.utm.my/id/eprint/88086/ In silico of the different pin size of unilateral tibia external fixator: a pilot study Zainudin, Nur Amalina Ramlee, Muhammad Hanif QH301 Biology External fixator is the promising medical device that can be used for treating bone fracture of tibia. The healing process can be taken place if some parameters are considered such as biomechanical aspect. One of them is the diameter of pin. However, the information regarding the biomechanics characteristics of the different pin size is fairly limited in the literature, thus it might be difficult for medical surgeons to choose pin size. Therefore, this pilot study was conducted to study the different pin size in treating transverse fracture in order to investigate the effect of it. Three-dimensional model of tibia bone and external fixator was constructed using MIMICS and Solid works, respectively. For the analysis, finite element method was utilized to simulate a stance phase condition where 50% of body weight was applied axially on the proximal tibia. The results showed that the 6.5 mm pin size demonstrated the lowest von mises stress at the pin bone interface. ECronicon 2019 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/88086/1/MuhammadHanifRamlee2019_InSilicooftheDifferentPinSize.pdf Zainudin, Nur Amalina and Ramlee, Muhammad Hanif (2019) In silico of the different pin size of unilateral tibia external fixator: a pilot study. EC Orthopaedics, 10 (11). pp. 7-10. https://www.ecronicon.com/ecor/pdf/ECOR-10-00512.pdf
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic QH301 Biology
spellingShingle QH301 Biology
Zainudin, Nur Amalina
Ramlee, Muhammad Hanif
In silico of the different pin size of unilateral tibia external fixator: a pilot study
description External fixator is the promising medical device that can be used for treating bone fracture of tibia. The healing process can be taken place if some parameters are considered such as biomechanical aspect. One of them is the diameter of pin. However, the information regarding the biomechanics characteristics of the different pin size is fairly limited in the literature, thus it might be difficult for medical surgeons to choose pin size. Therefore, this pilot study was conducted to study the different pin size in treating transverse fracture in order to investigate the effect of it. Three-dimensional model of tibia bone and external fixator was constructed using MIMICS and Solid works, respectively. For the analysis, finite element method was utilized to simulate a stance phase condition where 50% of body weight was applied axially on the proximal tibia. The results showed that the 6.5 mm pin size demonstrated the lowest von mises stress at the pin bone interface.
format Article
author Zainudin, Nur Amalina
Ramlee, Muhammad Hanif
author_facet Zainudin, Nur Amalina
Ramlee, Muhammad Hanif
author_sort Zainudin, Nur Amalina
title In silico of the different pin size of unilateral tibia external fixator: a pilot study
title_short In silico of the different pin size of unilateral tibia external fixator: a pilot study
title_full In silico of the different pin size of unilateral tibia external fixator: a pilot study
title_fullStr In silico of the different pin size of unilateral tibia external fixator: a pilot study
title_full_unstemmed In silico of the different pin size of unilateral tibia external fixator: a pilot study
title_sort in silico of the different pin size of unilateral tibia external fixator: a pilot study
publisher ECronicon
publishDate 2019
url http://eprints.utm.my/id/eprint/88086/1/MuhammadHanifRamlee2019_InSilicooftheDifferentPinSize.pdf
http://eprints.utm.my/id/eprint/88086/
https://www.ecronicon.com/ecor/pdf/ECOR-10-00512.pdf
_version_ 1687393524494893056