Computer based laboratory electrocardiogram measurements on the outpatients of the De La Salle University-Medical Center, Dasmarinas, Cavite
Computer based laboratory electrocardiogram (ECG) measurements using the PASCO equipment was conducted simultaneously with a KENZ ECG 106 on 9 outpatients of De La Salle University-Medical Center, Dasmarinas, Cavite. The QRS-wave time interval value of both equipment was compared and analyzed using...
Saved in:
Main Author: | |
---|---|
Format: | text |
Language: | English |
Published: |
Animo Repository
1998
|
Subjects: | |
Online Access: | https://animorepository.dlsu.edu.ph/etd_bachelors/1615 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | De La Salle University |
Language: | English |
Summary: | Computer based laboratory electrocardiogram (ECG) measurements using the PASCO equipment was conducted simultaneously with a KENZ ECG 106 on 9 outpatients of De La Salle University-Medical Center, Dasmarinas, Cavite. The QRS-wave time interval value of both equipment was compared and analyzed using Within-subject designs of the t-test statistical analysis. Results indicate that there were significant differences on the QRS-wave time interval values between the PASCO ECG QRS-wave readings and the Lead 1 KENZ ECG-106 QRS-wave readings. The said results prompted the study to conclude that the PASCO ECG QRS-wave time interval readings are not comparable to the QRS-wave time interval of the Lead of the KENZ ECG. This conclusion was supported by the PASCO ECG readings on the 7 patients that were tested for the study and were rejected by the t-test analysis. The remaining 3 patients were only accepted because there are not enough data to prove otherwise. |
---|