Multi-angle data acquisition to compensate transducer finite size in photoacoustic tomography

In photoacoustic tomography (PAT) systems, the tangential resolution decreases due to the finite size of the transducer as the off-center distance increases. To address this problem, we propose a multi-angle detection approach in which the transducer used for data acquisition rotates around its cent...

全面介紹

Saved in:
書目詳細資料
Main Authors: Hakakzadeh, Soheil, Mozaffarzadeh, Moein, Mostafavi, Seyed Masood, Kavehvash, Zahra, Rajendran, Praveenbalaji, Verweij, Martin, de Jong, Nico, Pramanik, Manojit
其他作者: School of Chemistry, Chemical Engineering and Biotechnology
格式: Article
語言:English
出版: 2022
主題:
在線閱讀:https://hdl.handle.net/10356/162727
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:In photoacoustic tomography (PAT) systems, the tangential resolution decreases due to the finite size of the transducer as the off-center distance increases. To address this problem, we propose a multi-angle detection approach in which the transducer used for data acquisition rotates around its center (with specific angles) as well as around the scanning center. The angles are calculated based on the central frequency and diameter of the transducer and the radius of the region-of-interest (ROI). Simulations with point-like absorbers (for point-spread-function evaluation) and a vasculature phantom (for quality assessment), and experiments with ten 0.5 mm-diameter pencil leads and a leaf skeleton phantom are used for evaluation of the proposed approach. The results show that a location-independent tangential resolution is achieved with 150 spatial sampling and central rotations with angles of ±8°/±16°. With further developments, the proposed detection strategy can replace the conventional detection (rotating a transducer around ROI) in PAT.