First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study

This report conducts a comprehensive phantom study on the dual-energy computed tomography (DECT) system of General Electric (GE) Revolution™ CT (GE Healthcare, Chicago, Illinois, USA), which will be used by National Cancer Centre Singapore (NCCS). This study is done with the use of a new commercial...

Full description

Saved in:
Bibliographic Details
Main Author: Low, Reece Cin
Other Authors: -
Format: Final Year Project
Language:English
Published: Nanyang Technological University 2023
Subjects:
Online Access:https://hdl.handle.net/10356/166248
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
id sg-ntu-dr.10356-166248
record_format dspace
spelling sg-ntu-dr.10356-1662482023-04-24T15:36:13Z First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study Low, Reece Cin - School of Physical and Mathematical Sciences National Cancer Centre Singapore Tan Hong Qi James Lee hqtan1988@gmail.com, jameslee@ntu.edu.sg Science::Physics::Radiation physics This report conducts a comprehensive phantom study on the dual-energy computed tomography (DECT) system of General Electric (GE) Revolution™ CT (GE Healthcare, Chicago, Illinois, USA), which will be used by National Cancer Centre Singapore (NCCS). This study is done with the use of a new commercial multi-energy computed tomography (MECT) phantom - Multi-Energy CT Phantom (Sun Nuclear, Model 1472, Melbourne, Florida, USA). Quality assurance (QA) of the DECT system is necessary to ensure that all images acquired are functioning optimally and can be trusted and used with confidence during treatment planning. However, there is currently no published guideline on how to best QA a DECT system. Hence, to QA the DECT system in this study, the functionality of two applications of DECT - monoenergetic reconstruction and ρ - Zeff decomposition - was evaluated. Results show that the HU and effective atomic number measurements obtained using the system were generally accurate. Findings also indicate that there is a dependence of HU accuracy on beam energy, concentration of iodine and calcium, as well as phantom size, and there is also a dependence of effective atomic number accuracy on fitting parameters used. However, since all the HU results obtained were within tolerance and there is a strong correlation between measured and calculated effective atomic numbers, the results obtained are considered acceptable and could be used in clinical practice. This validates the proper functioning of monoenergetic reconstruction and ρ - Zeff decomposition in the DECT system and signifies that the methods and measurement procedures employed in this study can be used in future routine QA procedures. Bachelor of Science in Physics 2023-04-24T07:01:09Z 2023-04-24T07:01:09Z 2023 Final Year Project (FYP) Low, R. C. (2023). First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study. Final Year Project (FYP), Nanyang Technological University, Singapore. https://hdl.handle.net/10356/166248 https://hdl.handle.net/10356/166248 en application/pdf Nanyang Technological University
institution Nanyang Technological University
building NTU Library
continent Asia
country Singapore
Singapore
content_provider NTU Library
collection DR-NTU
language English
topic Science::Physics::Radiation physics
spellingShingle Science::Physics::Radiation physics
Low, Reece Cin
First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
description This report conducts a comprehensive phantom study on the dual-energy computed tomography (DECT) system of General Electric (GE) Revolution™ CT (GE Healthcare, Chicago, Illinois, USA), which will be used by National Cancer Centre Singapore (NCCS). This study is done with the use of a new commercial multi-energy computed tomography (MECT) phantom - Multi-Energy CT Phantom (Sun Nuclear, Model 1472, Melbourne, Florida, USA). Quality assurance (QA) of the DECT system is necessary to ensure that all images acquired are functioning optimally and can be trusted and used with confidence during treatment planning. However, there is currently no published guideline on how to best QA a DECT system. Hence, to QA the DECT system in this study, the functionality of two applications of DECT - monoenergetic reconstruction and ρ - Zeff decomposition - was evaluated. Results show that the HU and effective atomic number measurements obtained using the system were generally accurate. Findings also indicate that there is a dependence of HU accuracy on beam energy, concentration of iodine and calcium, as well as phantom size, and there is also a dependence of effective atomic number accuracy on fitting parameters used. However, since all the HU results obtained were within tolerance and there is a strong correlation between measured and calculated effective atomic numbers, the results obtained are considered acceptable and could be used in clinical practice. This validates the proper functioning of monoenergetic reconstruction and ρ - Zeff decomposition in the DECT system and signifies that the methods and measurement procedures employed in this study can be used in future routine QA procedures.
author2 -
author_facet -
Low, Reece Cin
format Final Year Project
author Low, Reece Cin
author_sort Low, Reece Cin
title First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
title_short First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
title_full First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
title_fullStr First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
title_full_unstemmed First report of the quality assurance of a dual-energy computed tomography (DECT) system: a phantom study
title_sort first report of the quality assurance of a dual-energy computed tomography (dect) system: a phantom study
publisher Nanyang Technological University
publishDate 2023
url https://hdl.handle.net/10356/166248
_version_ 1764208022578528256