Design and evaluation of a light-field multi-wavelength pyrometer
This letter describes the design and implementation of a multi-wavelength light-field pyrometer, where six-channel radiation images were captured with one CMOS sensor. Such capability is achieved by placing a 2 × 3 filter array in front of the main lens of an unfocused light-field camera, such that...
Saved in:
Main Authors: | , , , , |
---|---|
Other Authors: | |
Format: | Article |
Language: | English |
Published: |
2023
|
Subjects: | |
Online Access: | https://hdl.handle.net/10356/170748 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Nanyang Technological University |
Language: | English |
Summary: | This letter describes the design and implementation of a multi-wavelength light-field pyrometer, where six-channel radiation images were captured with one CMOS sensor. Such capability is achieved by placing a 2 × 3 filter array in front of the main lens of an unfocused light-field camera, such that discrete wavelength and radiation intensity can be simultaneously recorded. It demonstrates that through black-body furnace experiments, how multi-channel radiation images can be extracted from one raw light-field multispectral image, and how accurate 2D temperature distribution can be recovered by optimization algorithms. |
---|