Experimental investigation of leakage flow in centrifugal blood pump

The study of leakage flow is crucial in designing centrifugal blood pumps. High volumetric efficiency is achieved when the leakage flow is low. However, low leakage flow cause formation of thrombus, while too high a leakage flow causes high shear to blood resulting in hemolysis. In this proje...

Full description

Saved in:
Bibliographic Details
Main Author: Aw, Chung Seng
Other Authors: Chan Weng Kong
Format: Final Year Project
Language:English
Published: 2012
Subjects:
Online Access:http://hdl.handle.net/10356/50188
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
id sg-ntu-dr.10356-50188
record_format dspace
spelling sg-ntu-dr.10356-501882023-03-04T19:07:53Z Experimental investigation of leakage flow in centrifugal blood pump Aw, Chung Seng Chan Weng Kong Wong Yew Wah School of Mechanical and Aerospace Engineering DRNTU::Engineering::Mechanical engineering::Fluid mechanics The study of leakage flow is crucial in designing centrifugal blood pumps. High volumetric efficiency is achieved when the leakage flow is low. However, low leakage flow cause formation of thrombus, while too high a leakage flow causes high shear to blood resulting in hemolysis. In this project, designing and fabrication of dummy impellers was first carried out. The blood pump test rig was checked and being serviced to ensure good working condition. Experiments were first conducted to establish the pump characteristics of the centrifugal pump with dummy impellers. Also, the ΔP of outer and inner radii of the volute was measured. Generally, the pump characteristics curves obtained show that pump head increased with factors like throttling of flow rate, increase in pump speed increases and increase in gap size. From the results, the highest gap size h = 0.50 mm has best performance. However, the magnitude of increase in performance was found to decrease with increasing gap size. Changing the gap size, pump speed and throttling of flow did affect the ΔP distribution similar to the overall pump characteristics curves. The ΔP at rear volute is lower than the front volute, at lower operating flow rate. At higher flow rate, the ΔP at rear volute higher. This is due to the difference in geometry of front and rear volute. Another set of experiments was performed to obtain leakage flow measurement of the centrifugal blood pump using auxiliary pumps to provide suitable operating conditions such that reverse flow could be observed. First, a two-pump in series was introduced; results showed that the auxiliary pump were not strong enough to create high ΔP measurements at the volute face. Then, it was modified to a four-pump in series; results showed that at the lowest gap size, h = 0.225 mm, the ΔP was found to have of the same range of pressure difference with the previous experimental result. Leakage flow of 2.8 -3.8 l/min would be expected if the pump has the same geometry and operating at the same speed of 2000 rpm. The leakage flow result was higher compared to the previous result. The results obtained were compared to the theoretical model. It was found that the relationship of ΔP with corresponding leakage flow rate did not behave linearly unlike the theoretical model proposed. For higher gap size, h = 0.40 mm, the experimental data was found to be close to the theoretical result. Finally, the results were used to compare to the previous CFX result. The result showed that the experimental leakage flow was much higher than the CFX result for all the gap sizes and was in agreement of each other. Bachelor of Engineering (Mechanical Engineering) 2012-05-30T09:12:53Z 2012-05-30T09:12:53Z 2012 2012 Final Year Project (FYP) http://hdl.handle.net/10356/50188 en Nanyang Technological University 87 p. application/pdf
institution Nanyang Technological University
building NTU Library
continent Asia
country Singapore
Singapore
content_provider NTU Library
collection DR-NTU
language English
topic DRNTU::Engineering::Mechanical engineering::Fluid mechanics
spellingShingle DRNTU::Engineering::Mechanical engineering::Fluid mechanics
Aw, Chung Seng
Experimental investigation of leakage flow in centrifugal blood pump
description The study of leakage flow is crucial in designing centrifugal blood pumps. High volumetric efficiency is achieved when the leakage flow is low. However, low leakage flow cause formation of thrombus, while too high a leakage flow causes high shear to blood resulting in hemolysis. In this project, designing and fabrication of dummy impellers was first carried out. The blood pump test rig was checked and being serviced to ensure good working condition. Experiments were first conducted to establish the pump characteristics of the centrifugal pump with dummy impellers. Also, the ΔP of outer and inner radii of the volute was measured. Generally, the pump characteristics curves obtained show that pump head increased with factors like throttling of flow rate, increase in pump speed increases and increase in gap size. From the results, the highest gap size h = 0.50 mm has best performance. However, the magnitude of increase in performance was found to decrease with increasing gap size. Changing the gap size, pump speed and throttling of flow did affect the ΔP distribution similar to the overall pump characteristics curves. The ΔP at rear volute is lower than the front volute, at lower operating flow rate. At higher flow rate, the ΔP at rear volute higher. This is due to the difference in geometry of front and rear volute. Another set of experiments was performed to obtain leakage flow measurement of the centrifugal blood pump using auxiliary pumps to provide suitable operating conditions such that reverse flow could be observed. First, a two-pump in series was introduced; results showed that the auxiliary pump were not strong enough to create high ΔP measurements at the volute face. Then, it was modified to a four-pump in series; results showed that at the lowest gap size, h = 0.225 mm, the ΔP was found to have of the same range of pressure difference with the previous experimental result. Leakage flow of 2.8 -3.8 l/min would be expected if the pump has the same geometry and operating at the same speed of 2000 rpm. The leakage flow result was higher compared to the previous result. The results obtained were compared to the theoretical model. It was found that the relationship of ΔP with corresponding leakage flow rate did not behave linearly unlike the theoretical model proposed. For higher gap size, h = 0.40 mm, the experimental data was found to be close to the theoretical result. Finally, the results were used to compare to the previous CFX result. The result showed that the experimental leakage flow was much higher than the CFX result for all the gap sizes and was in agreement of each other.
author2 Chan Weng Kong
author_facet Chan Weng Kong
Aw, Chung Seng
format Final Year Project
author Aw, Chung Seng
author_sort Aw, Chung Seng
title Experimental investigation of leakage flow in centrifugal blood pump
title_short Experimental investigation of leakage flow in centrifugal blood pump
title_full Experimental investigation of leakage flow in centrifugal blood pump
title_fullStr Experimental investigation of leakage flow in centrifugal blood pump
title_full_unstemmed Experimental investigation of leakage flow in centrifugal blood pump
title_sort experimental investigation of leakage flow in centrifugal blood pump
publishDate 2012
url http://hdl.handle.net/10356/50188
_version_ 1759853561547587584