Flow investigation in a centrifugal pump

Centrifugal blood pumps are becoming the popular choice for VADs because of their simple design and low costs. Nevertheless, the presence of flow stagnation or excessive leakage flow would induce thrombosis and hemolysis. Since these risks are flow related, ANSYS Workbench v15.0 was employed for the...

Full description

Saved in:
Bibliographic Details
Main Author: Ang, Aaron Guo Hao
Other Authors: Chan Weng Kong
Format: Final Year Project
Language:English
Published: 2015
Subjects:
Online Access:http://hdl.handle.net/10356/64479
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
id sg-ntu-dr.10356-64479
record_format dspace
spelling sg-ntu-dr.10356-644792023-03-04T18:41:12Z Flow investigation in a centrifugal pump Ang, Aaron Guo Hao Chan Weng Kong Wong Yew Wah School of Mechanical and Aerospace Engineering DRNTU::Engineering::Mechanical engineering Centrifugal blood pumps are becoming the popular choice for VADs because of their simple design and low costs. Nevertheless, the presence of flow stagnation or excessive leakage flow would induce thrombosis and hemolysis. Since these risks are flow related, ANSYS Workbench v15.0 was employed for the CFD simulations to investigate flow characteristics in the pump. CFD simulations of the pump were performed for four gap widths (0.2mm, 0.3mm, 0.4mm and 0.5mm) at operating flow rates of two to six liters per minute for impeller speed of 2000rpm. It was found that pump head, P increases as operating flow rate, L and gap width decrease. Additionally, total leakage flow rate, QL increases along with differential pressure in the gap, ΔP and the gap width but decreases as the operating flow rate, Q escalates. These trends are congruent with available experimental results. The only difference is that the experiments obtained QL values from experimental trials of a pump with a dummy. Generally, the maximum QL discrepancy between the CFD simulations and experiments does not exceed 15.7% except for the 0.5mm gap pump, making it difficult to ascertain the correlation of the CFD and experimental results. Consequently, CFD simulations of the pump with a dummy for the similar four gap widths at dummy speed of 2000rpm were carried out. The maximum difference in QLvalues between simulations of the pump and the pump with a dummy is 3.4% for the 0.2mm and 0.3mm gap widths and 15.9% for the 0.4mm and 0.5mm gap widths. Thus, using the pump with a dummy to predict QL values is relatively reliable. Comparing the pump with a dummy between the CFD simulations and the experiments, CFD simulations consistently under predict the ΔP values for a given QL due to its inability to accurately predict the losses. However, it over predicted the ΔP values for the 0.5mm gap pump with a dummy. This abnormally is related back to and accounts for the disparities in QL values between CFD simulations and the experiments, justifying the suspicion that there are systematic errors when the experiment for the 0.5mm gap pump with a dummy was conducted. Concluding, the CFD software predicts the leakage flow rates quite accurately. Bachelor of Engineering (Mechanical Engineering) 2015-05-27T03:39:49Z 2015-05-27T03:39:49Z 2015 2015 Final Year Project (FYP) http://hdl.handle.net/10356/64479 en Nanyang Technological University 92 p. application/pdf
institution Nanyang Technological University
building NTU Library
continent Asia
country Singapore
Singapore
content_provider NTU Library
collection DR-NTU
language English
topic DRNTU::Engineering::Mechanical engineering
spellingShingle DRNTU::Engineering::Mechanical engineering
Ang, Aaron Guo Hao
Flow investigation in a centrifugal pump
description Centrifugal blood pumps are becoming the popular choice for VADs because of their simple design and low costs. Nevertheless, the presence of flow stagnation or excessive leakage flow would induce thrombosis and hemolysis. Since these risks are flow related, ANSYS Workbench v15.0 was employed for the CFD simulations to investigate flow characteristics in the pump. CFD simulations of the pump were performed for four gap widths (0.2mm, 0.3mm, 0.4mm and 0.5mm) at operating flow rates of two to six liters per minute for impeller speed of 2000rpm. It was found that pump head, P increases as operating flow rate, L and gap width decrease. Additionally, total leakage flow rate, QL increases along with differential pressure in the gap, ΔP and the gap width but decreases as the operating flow rate, Q escalates. These trends are congruent with available experimental results. The only difference is that the experiments obtained QL values from experimental trials of a pump with a dummy. Generally, the maximum QL discrepancy between the CFD simulations and experiments does not exceed 15.7% except for the 0.5mm gap pump, making it difficult to ascertain the correlation of the CFD and experimental results. Consequently, CFD simulations of the pump with a dummy for the similar four gap widths at dummy speed of 2000rpm were carried out. The maximum difference in QLvalues between simulations of the pump and the pump with a dummy is 3.4% for the 0.2mm and 0.3mm gap widths and 15.9% for the 0.4mm and 0.5mm gap widths. Thus, using the pump with a dummy to predict QL values is relatively reliable. Comparing the pump with a dummy between the CFD simulations and the experiments, CFD simulations consistently under predict the ΔP values for a given QL due to its inability to accurately predict the losses. However, it over predicted the ΔP values for the 0.5mm gap pump with a dummy. This abnormally is related back to and accounts for the disparities in QL values between CFD simulations and the experiments, justifying the suspicion that there are systematic errors when the experiment for the 0.5mm gap pump with a dummy was conducted. Concluding, the CFD software predicts the leakage flow rates quite accurately.
author2 Chan Weng Kong
author_facet Chan Weng Kong
Ang, Aaron Guo Hao
format Final Year Project
author Ang, Aaron Guo Hao
author_sort Ang, Aaron Guo Hao
title Flow investigation in a centrifugal pump
title_short Flow investigation in a centrifugal pump
title_full Flow investigation in a centrifugal pump
title_fullStr Flow investigation in a centrifugal pump
title_full_unstemmed Flow investigation in a centrifugal pump
title_sort flow investigation in a centrifugal pump
publishDate 2015
url http://hdl.handle.net/10356/64479
_version_ 1759857491035815936