Experimental investigation of a bi-leaflet heart valve
Babies with congenital heart defects require high risk surgeries at birth or adolescence. As the patient matures, replacement of larger artificial valves to accommodate their growth would require additional surgeries. The objective of this experiment is to reduce the amount of re-surgeries required...
Saved in:
主要作者: | |
---|---|
其他作者: | |
格式: | Final Year Project |
語言: | English |
出版: |
2016
|
主題: | |
在線閱讀: | http://hdl.handle.net/10356/68518 |
標簽: |
添加標簽
沒有標簽, 成為第一個標記此記錄!
|
機構: | Nanyang Technological University |
語言: | English |
總結: | Babies with congenital heart defects require high risk surgeries at birth or adolescence. As the patient matures, replacement of larger artificial valves to accommodate their growth would require additional surgeries. The objective of this experiment is to reduce the amount of re-surgeries required by determining the compatibly of the artificial valves upon the change in diameters of the artery with growth and the method of which the valve was sutured.
The valves used are made of expanded polytetrafluoroethylene (ePTFE) with a 13mm diameter kept constant for the various diameters of the silicon conduits ranging from 13mm, 17mm and 21mm.
From the experimental results of the 3 different diameters of silicon conduits, the percentage of regurgitation ranged from 7% to 76%.The results showed that the increase in percentage of regurgitation increases linearly with the increase in the diameter of the lumen. Methods used in suturing of the valve affected the regurgitation rate was proven with the experimental results.
In comparison to previous experimental results by Jowell Cheng, a correlation found between the 13-21mm silicon conduits yield a 10% variation which can be attributed to different experimental conditions. |
---|