Multiple single-element transducer photoacoustic computed tomography system
Light absorption by the chromophores (hemoglobin, melanin, water etc.) present in any biological tissue results in local temperature rise. This rise in temperature results in generation of pressure waves due to the thermoelastic expansion of the tissue. In a circular scanning photoacoustic computed...
Saved in:
Main Authors: | , , |
---|---|
Other Authors: | |
Format: | Conference or Workshop Item |
Language: | English |
Published: |
2018
|
Subjects: | |
Online Access: | https://hdl.handle.net/10356/87512 http://hdl.handle.net/10220/44468 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Nanyang Technological University |
Language: | English |
id |
sg-ntu-dr.10356-87512 |
---|---|
record_format |
dspace |
spelling |
sg-ntu-dr.10356-875122023-12-29T06:44:14Z Multiple single-element transducer photoacoustic computed tomography system Kalva, Sandeep Kumar Hui, Zhe Zhi Pramanik, Manojit School of Chemical and Biomedical Engineering Proceedings of SPIE - Photons Plus Ultrasound: Imaging and Sensing 2018 Single Element Ultrasound Transducer Multiple Ultrasound Transducers Light absorption by the chromophores (hemoglobin, melanin, water etc.) present in any biological tissue results in local temperature rise. This rise in temperature results in generation of pressure waves due to the thermoelastic expansion of the tissue. In a circular scanning photoacoustic computed tomography (PACT) system, these pressure waves can be detected using a single-element ultrasound transducer (SUST) (while rotating in full 360° around the sample) or using a circular array transducer. SUST takes several minutes to acquire the PA data around the sample whereas the circular array transducer takes only a fraction of seconds. Hence, for real time imaging circular array transducers are preferred. However, these circular array transducers are custom made, expensive and not easily available in the market whereas SUSTs are cheap and readily available in the market. Using SUST for PACT systems is still cost effective. In order to reduce the scanning time to few seconds instead of using single SUST (rotating 360° ), multiple SUSTs can be used at the same time to acquire the PA data. This will reduce the scanning time by two-fold in case of two SUSTs (rotating 180° ) or by four-fold and eight-fold in case of four SUSTs (rotating 90° ) and eight SUSTs (rotating 45° ) respectively. Here we show that with multiple SUSTs, similar PA images (numerical and experimental phantom data) can be obtained as that of PA images obtained using single SUST. Published version 2018-02-28T08:42:31Z 2019-12-06T16:43:28Z 2018-02-28T08:42:31Z 2019-12-06T16:43:28Z 2018-02-01 2018 Conference Paper Kalva, S. K., Hui, Z. Z., & Pramanik, M. (2018). Multiple single-element transducer photoacoustic computed tomography system. Proceedings of SPIE - Photons Plus Ultrasound: Imaging and Sensing 2018, 10494, 104944D-. 203397 https://hdl.handle.net/10356/87512 http://hdl.handle.net/10220/44468 10.1117/12.2286606 203397 en © 2018 Society of Photo-optical Instrumentation Engineers (SPIE). This paper was published in Proceedings of SPIE - Photons Plus Ultrasound: Imaging and Sensing 2018 and is made available as an electronic reprint (preprint) with permission of SPIE. The published version is available at: [http://dx.doi.org/10.1117/12.2286606]. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper is prohibited and is subject to penalties under law. 7 p. application/pdf |
institution |
Nanyang Technological University |
building |
NTU Library |
continent |
Asia |
country |
Singapore Singapore |
content_provider |
NTU Library |
collection |
DR-NTU |
language |
English |
topic |
Single Element Ultrasound Transducer Multiple Ultrasound Transducers |
spellingShingle |
Single Element Ultrasound Transducer Multiple Ultrasound Transducers Kalva, Sandeep Kumar Hui, Zhe Zhi Pramanik, Manojit Multiple single-element transducer photoacoustic computed tomography system |
description |
Light absorption by the chromophores (hemoglobin, melanin, water etc.) present in any biological tissue results in local temperature rise. This rise in temperature results in generation of pressure waves due to the thermoelastic expansion of the tissue. In a circular scanning photoacoustic computed tomography (PACT) system, these pressure waves can be detected using a single-element ultrasound transducer (SUST) (while rotating in full 360° around the sample) or using a circular array transducer. SUST takes several minutes to acquire the PA data around the sample whereas the circular array transducer takes only a fraction of seconds. Hence, for real time imaging circular array transducers are preferred. However, these circular array transducers are custom made, expensive and not easily available in the market whereas SUSTs are cheap and readily available in the market. Using SUST for PACT systems is still cost effective. In order to reduce the scanning time to few seconds instead of using single SUST (rotating 360° ), multiple SUSTs can be used at the same time to acquire the PA data. This will reduce the scanning time by two-fold in case of two SUSTs (rotating 180° ) or by four-fold and eight-fold in case of four SUSTs (rotating 90° ) and eight SUSTs (rotating 45° ) respectively. Here we show that with multiple SUSTs, similar PA images (numerical and experimental phantom data) can be obtained as that of PA images obtained using single SUST. |
author2 |
School of Chemical and Biomedical Engineering |
author_facet |
School of Chemical and Biomedical Engineering Kalva, Sandeep Kumar Hui, Zhe Zhi Pramanik, Manojit |
format |
Conference or Workshop Item |
author |
Kalva, Sandeep Kumar Hui, Zhe Zhi Pramanik, Manojit |
author_sort |
Kalva, Sandeep Kumar |
title |
Multiple single-element transducer photoacoustic computed tomography system |
title_short |
Multiple single-element transducer photoacoustic computed tomography system |
title_full |
Multiple single-element transducer photoacoustic computed tomography system |
title_fullStr |
Multiple single-element transducer photoacoustic computed tomography system |
title_full_unstemmed |
Multiple single-element transducer photoacoustic computed tomography system |
title_sort |
multiple single-element transducer photoacoustic computed tomography system |
publishDate |
2018 |
url |
https://hdl.handle.net/10356/87512 http://hdl.handle.net/10220/44468 |
_version_ |
1787136753004445696 |