High resolution and deep tissue imaging using a near infrared acoustic resolution photoacoustic microscopy

It is always a great challenge for pure optical techniques to maintain good resolution and imaging depth at the same time. Photoacoustic imaging is an emerging technique which can overcome the limitation by pulsed light illumination and acoustic detection. Here, we report a Near Infrared Acoustic-Re...

Full description

Saved in:
Bibliographic Details
Main Authors: Moothanchery, Mohesh, Sharma, Arunima, Periyasamy, Vijitha, Pramanik, Manojit
Other Authors: School of Chemical and Biomedical Engineering
Format: Conference or Workshop Item
Language:English
Published: 2018
Subjects:
PAM
Online Access:https://hdl.handle.net/10356/87991
http://hdl.handle.net/10220/44471
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
Description
Summary:It is always a great challenge for pure optical techniques to maintain good resolution and imaging depth at the same time. Photoacoustic imaging is an emerging technique which can overcome the limitation by pulsed light illumination and acoustic detection. Here, we report a Near Infrared Acoustic-Resolution Photoacoustic Microscopy (NIR-AR-PAM) systm with 30 MHz transducer and 1064 nm illumination which can achieve a lateral resolution of around 88 μm and imaging depth of 9.2 mm. Compared to visible light NIR beam can penetrate deeper in biological tissue due to weaker optical attenuation. In this work, we also demonstrated the in vivo imaging capabilty of NIRARPAM by near infrared detection of SLN with black ink as exogenous photoacoustic contrast agent in a rodent model.