A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples

The 'Hemocue' device for rapid estimation of haemoglobin concentration was evaluated in a clinical setting. Repeatable accuracy of capillary, venous and arterial samples was examined and then compared with standard laboratory venous haemoglobin estimates using a 'Coulter JT' anal...

Full description

Saved in:
Bibliographic Details
Main Authors: Chen, P. P., Short, T. G., Leung, Denis H. Y., Oh, T. E.
Format: text
Language:English
Published: Institutional Knowledge at Singapore Management University 1992
Subjects:
Online Access:https://ink.library.smu.edu.sg/soe_research/294
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Singapore Management University
Language: English
id sg-smu-ink.soe_research-1293
record_format dspace
spelling sg-smu-ink.soe_research-12932010-09-23T05:48:03Z A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples Chen, P. P. Short, T. G. Leung, Denis H. Y. Oh, T. E. The 'Hemocue' device for rapid estimation of haemoglobin concentration was evaluated in a clinical setting. Repeatable accuracy of capillary, venous and arterial samples was examined and then compared with standard laboratory venous haemoglobin estimates using a 'Coulter JT' analyser in 42 patients. The mean values for haemoglobin (g/l) and coefficient of variation were capillary 108.2 (8.0); venous 104.9 (2.2); arterial 105.9 (2.0); and laboratory venous 104.6 (1.3). Although the mean haemoglobin values were similar, capillary samples were significantly less repeatable than venous or arterial samples (Pitman test, P < 0.001). Comparison of variance between the laboratory sample and each sampling technique demonstrated that capillary samples were significantly more variable than venous or arterial samples. Peripheral skin temperature did not influence the accuracy of capillary samples. Hemocue estimations of venous samples were found to be as accurate as laboratory estimations. The lack of repeatable accuracy of capillary estimations was sufficiently large that their use cannot be recommended in clinical practice. 1992-01-01T08:00:00Z text https://ink.library.smu.edu.sg/soe_research/294 info:doi/10.1177/0310057X9202000419 Research Collection School Of Economics eng Institutional Knowledge at Singapore Management University Econometrics Medicine and Health Sciences
institution Singapore Management University
building SMU Libraries
continent Asia
country Singapore
Singapore
content_provider SMU Libraries
collection InK@SMU
language English
topic Econometrics
Medicine and Health Sciences
spellingShingle Econometrics
Medicine and Health Sciences
Chen, P. P.
Short, T. G.
Leung, Denis H. Y.
Oh, T. E.
A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
description The 'Hemocue' device for rapid estimation of haemoglobin concentration was evaluated in a clinical setting. Repeatable accuracy of capillary, venous and arterial samples was examined and then compared with standard laboratory venous haemoglobin estimates using a 'Coulter JT' analyser in 42 patients. The mean values for haemoglobin (g/l) and coefficient of variation were capillary 108.2 (8.0); venous 104.9 (2.2); arterial 105.9 (2.0); and laboratory venous 104.6 (1.3). Although the mean haemoglobin values were similar, capillary samples were significantly less repeatable than venous or arterial samples (Pitman test, P < 0.001). Comparison of variance between the laboratory sample and each sampling technique demonstrated that capillary samples were significantly more variable than venous or arterial samples. Peripheral skin temperature did not influence the accuracy of capillary samples. Hemocue estimations of venous samples were found to be as accurate as laboratory estimations. The lack of repeatable accuracy of capillary estimations was sufficiently large that their use cannot be recommended in clinical practice.
format text
author Chen, P. P.
Short, T. G.
Leung, Denis H. Y.
Oh, T. E.
author_facet Chen, P. P.
Short, T. G.
Leung, Denis H. Y.
Oh, T. E.
author_sort Chen, P. P.
title A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
title_short A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
title_full A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
title_fullStr A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
title_full_unstemmed A Clinical Evaluation of the Hemocue Haemoglobinometer Using Capillary, Venous and Arterial Samples
title_sort clinical evaluation of the hemocue haemoglobinometer using capillary, venous and arterial samples
publisher Institutional Knowledge at Singapore Management University
publishDate 1992
url https://ink.library.smu.edu.sg/soe_research/294
_version_ 1770569101776781312