Kinking of catheters during translaryngeal jet ventilation: A bench model investigation of eight devices
Objective: To evaluate oxygen flow through several transtracheal devices in native and right angle kinked states. Material and Method: Eight catheter-over-needle, and two oxygen conveyance devices (Enk Flow Modulator 10L/min flow and Manujet III Jet device 15, 30, 50 psi) were examined. Oxygen flow...