A Multiple processing hybrid flow system for analysis of formaldehyde contamination in food
This work proposes a flow system suitable for the rapid screening of formaldehyde contaminated in food. The system is based on the concept of a flow analyzer with a Hantzsch reaction. An operating procedure was developed for multiple tasking and high sample throughput. This resulted in a significant...
Saved in:
Main Authors: | , , , , , , |
---|---|
Format: | Journal |
Published: |
2018
|
Subjects: | |
Online Access: | https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=77955723296&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/50675 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Chiang Mai University |
Summary: | This work proposes a flow system suitable for the rapid screening of formaldehyde contaminated in food. The system is based on the concept of a flow analyzer with a Hantzsch reaction. An operating procedure was developed for multiple tasking and high sample throughput. This resulted in a significant sample throughput of 51 samples h-1. Under the optimized conditions, linear calibration from 10 to 100 μM was obtained. The system gave a limit of detection and a limit of quantitation of 0.06 and 0.10 mg kg-1, respectively. The system was successfully applied to re-hydrated dry squids, vegetables and mushrooms. 2010 © The Japan Society for Analytical Chemistry. |
---|